Chapter 9 Probes
9-33
EV4-9 Biopsy Guide
Specication
Model Name
BP-KIT-024
Components
Biopsy Adapter 1 EA
Sheath 2 EA
Material
Stainless
Available Gauge
16G
Installation
1. After lling a sheath with ultrasound gel and covering the probe with it, tie the base of the probe
with a rubber band.
2. Mount the adapter onto the probe.
3. After covering the probe with a second sheath, tie the base of the probe so that the second tie
is lower than the rst one.
4. Insert the needle into the needle guide.