Operation Manual
4-42
Tips!
Threshold
This option allows the user to adjust the threshold value in order to eliminate unnecessary data
from images. As the number increases, cyst elements become more apparent. As the number
decreases bone elements become more apparent.
Orientation
Select 0, 90, 180, or 270° with the dial-button. The coordinate system rotates by the Z-axis in the “3D
standard coordinate system.
NOTE: This is available when Display is set to 2D/3D, ROI 3D or Fixed 3D.
U/D Flip
Each time you press the dial-button, the image is ipped upside down.
MPR Mode Menu
Display
Set the display method for 3D images.
2D
Images in the Axial, Sagittal and Coronal planes will be displayed.
Tips!
Since diagnosis is performed with multi-planar images only in 2D Mode, Full in the soft menu can
be used for more detailed observation.
2D / 3D
Images in the Axial, Sagittal or Coronal plane and 3D images will be displayed.
ROI 3D
This is the default mode when 3D View is enabled. Images in the Axial, Sagittal and Coronal planes
and 3D images will be displayed. Adjust the ROI box for observation.