Operation Manual
9-24
Cleaning and Disinfecting Plastic Biopsy Kits
■
Cleaning
1. After use, remove the biopsy kit from the probe.
2. Disassemble the biopsy kit into its components, if applicable. Discard the single-use parts. These
parts cannot be disinfected.
3. Using a small brush and water, scrub each part to remove trapped material from the re-usable
components.
4. Rinse with water to remove remaining particulates.
■
Disinfecting
CAUTION: A plastic biopsy kit can only be disinfected using chemically compatible cold
disinfectants. Disinfecting through autoclaving or by using gas or radiation will cause damage to it.
Please refer to the corresponding user manual for storage, use, and disposal of the disinfectant.
1. Check the disinfection duration (generally 10 hours) and temperature of the disinfectant.
2. After disinfection, follow the proper post-disinfection procedure for the disinfection method
used.
3. Inspect the components for damage such as cracks, rust or breakage. If any damage is found,
discontinue use of the biopsy kit and contact Samsung Medison Customer Service for assistance.