Chapter 9 Probes
9-11
Electric Shock
The probe uses electrical energy. When it touches conductive materials, there is a risk of electric shock
to the patient or the user.
WARNING:
X
Ensure inspection for potential short circuit from Samsung Medison Customer Service on a
regular basis.
X
Do not immerse the probe in liquid.
X
Do not drop the probe or apply mechanical shock.
X
Inspect the housing, strain relief, lens and seal for damage, and check for functional problems
before and after each use.
X
Do not apply excessive force to twist, pull or bend the probe cable. It may result in short circuit.
X
The power protection fuse protects the probe and the product from excess current. If the
power-monitoring protection circuit detects any excess current, it shuts off the current to the
equipment, preventing overheating and limiting the ultrasound power output.
X
For this product, the patient contact temperature is limited to 43°C and the ultrasound power
output (AP&I) should conform to U.S. FDA standards.