Chapter 9 Probes
9-23
Cleaning and Disinfecting Biopsy Kits
Clean and disinfect the biopsy kit to reduce pathogens to a level of 10-6. Some components of the biopsy
kit may be disposable. Please read the biopsy kit user manual carefully before use.
WARNING: Always use protective eyewear and gloves when cleaning and disinfecting a biopsy kit.
Cleaning and Disinfecting Stainless Steel Biopsy Kits
■
Cleaning
1. After use, remove the biopsy kit from the probe.
2. Disassemble the biopsy kit into its components, if applicable.
3. Using a small brush and water, scrub each part to remove trapped material from the biopsy kit.
4. Rinse with water to remove remaining particulates.
■
Disinfecting
1. Disinfect the adapter by autoclaving (steam) or by using gas (Ethylene Oxide).
2. After disinfection, follow the proper post-disinfection procedure for the disinfection method
used. (Please refer to the user manual for the disinfectant.)
3. Inspect the components for damage such as cracks, rust or breakage. If any damage is found,
discontinue use of the biopsy kit and contact Samsung Medison Customer Service for assistance.