Vista de página 93
4-8
User Manual
Starting 2D Mode
NOTE: As 2D Mode is included in all diagnosis modes, it cannot be closed.
Press the 2D button on the control panel. Pressing the 2D button in other modes will take you to 2D
Mode, which is the basic diagnosis mode.
2D Mode Menu
Following items are displayed on menu when you press the Menu dial-button in 2D mode. Select the
item and adjust the value with the Menu dial-button.
FA
FA stands for Frame Average.
When an image is updated, the current and previous images are averaged. Speckles may appear
when the same diagnosis area is scanned continuously. This function is used to minimize such
speckles. Select a value between 0- 9 with the Menu dial-button.
DR
DR stands for Dynamic Range.
Adjust contrast by changing the ratio of minimum/maximum input signal values. Select a value
between 40 - 180 with the Menu dial-button. Increasing the ratio renders softer images.
EE
EE stands for Edge Enhance.
This emphasizes the edges of tissues or organs. Select a value between 0 - 9 with the Menu dial-
button. Increase the Edge Enhance value to obtain clearer pictures of boundaries.
Reject
This function is used to eliminate noise or low level echoes for a clearer view of images. Select a
value between 1 - 32 with the Menu dial-button.