Reference Manual 145
EVN4-9: Pulsed Doppler & 2D+Pulsed Doppler
Index Label MI Scan
TIS TIB
TICNon-scan Non-scan
Non-scan
Aaprt≤1cm² Aaprt>1cm²
Maximum index value 1.400 0.594 0.617 0.483** 1.420 (b)
Pr.3 2.940
W0
2D P1:2.190
PD P1x1:21.920
PD P1x1:25.664 PD P:19.300 (b)
min[W.3,Ita.3] 20.105
z1 0.700
Zbp 1.043
Zsp 0.600 0.700
deq (zsp) 0.239
fc 4.420
2D :5.620
PD :5.130
5.050 5.050 5.160 (b)
X Aaprt
2D :1.730
PD :0.633
0.633 0.633 0.295 (b)
Y Aaprt 0.600 0.600 0.600 0.600 (b)
PD 0.829
PRF 999.000
Pr_Pii_max 3.020
deq_Pii_max 0.209
FLx (cm) 4.000 4.000 4.000 (b)
FLy (cm) 2.300 2.300 2.300 (b)
Ipa.MI_max 309.000
Control 1 MI
Control 2 TIS_as
Control 3 TIS_as_U TIS_bs
Control 4 TIB_bs
Control 5 TIC_as
Control 1: PulsedDoppler Mode, PD:4.4MHz,Focus:2.0Cm,Data No:3109721336-563195
Control 2: 2D+PulsedDoppler Mode, 2D:6.2MHzPD:5.1MHz,Focus:4.0Cm,FR:20.300,Data No:3110253259-717378
Control 3: PulsedDoppler Mode, PD:5.1MHz,Focus:4.0Cm,Data No:3109750355-515051
Control 4: PulsedDoppler Mode, PD:5.1MHz,Focus:2.0Cm,Data No:3109757202-619709
Control 5: PulsedDoppler Mode, PD:5.1MHz,Focus:1.0Cm,Data No:3109756048-259922
Patient-Applied Part Temperature Table
Transducer
Surface Temperature
In-Air (°C)
Surface Temperature
Simulated-Use (°C)
C2-4/20
<41 <41
C2-5
42.2 <41
C2-8
<41 <41
CN2-8
42.3 <41
CN4-9
<41 <41
EC4-9
<41 41.9
EVN4-9
<41 41.9
L5-12/60
<41 42.4
LE5-12
45.1 41.2
LN5-12
41.1 <41
LN5-12/40
<41 42.3