2-14
User Manual
Power Connection Part
The power connection part is located at the back of the system.
1
6
2
3
5
4
[Figure 2.7 Power Connection Part]
Item Descriptions
1
Power switch: Supplies or blocks power to the entire system.
2
Power inlet: For the power cable to connect to external power
3
Fuse holder: For the inlet fuse
4
Equipotential terminal: Should be connected to the equipotential bonding network in medical rooms.
5
INPUT/OUTPUT Voltage Selector: Selects the input/output power.
6
Auxiliary socket outlet: Supplies an external peripheral device with power from the product's internal
power supply
CAUTION:
X
Pressing the On/O button for longer than five seconds will immediately turn the power off
and result in damage to the hard disk. Please refrain from such use unless there is a serious
emergency.
X
To ensure that the product is safely cut off from electrical power, set the power switch at the
rear of the product to Off position after using the product.