8-5
Chapter 8 Maintenance
Disinfections
CAUTION: Use only recommended disinfectants on system surfaces.
A disinfectant qualied by the FDA 510(k) process is recommended. The following disinfectants are
recommended because of both their biological eectiveness (as qualied through the FDA 510(k)
process) and their chemical compatibility with Samsung Medison ultrasound products.
Solutions Country
Type Active ingredient FDA 510(k)
Cidex USA Liquid Gluteraldehyde K934434
Cidex Plus USA Liquid Gluteraldehyde K923744
1. Turn o the system and disconnect the system power cord from the wall outlet.
2. Mix the disinfection solution compatible with your system according to label instructions for
solution strength.
3. Wipe the system surfaces with the disinfectant solution, following the disinfectant label instructions
for wipe durations, solution strength, and disinfectant contact duration.
4. Air dry or towel dry with a sterile cloth according to the instructions on the disinfectant label.
Vista de página 258