Reference Manual 110
P
r
at max. I
pi
the PEAK-RAREFACTIONAL ACOUSTIC PRESSURE at the point where the free -eld
(non-attenuated), PULSE INTENSITY INTEGRAL is a maximum (megapascals).
prr the PULSE REPETITION RATE associated with the transmit pattern giving rise to
the reported value of MI (pulses per second).
TIB the BONE THERMAL INDEX for applications, such as fetal (second and third
trimester) or neonatal cephalic (through the fontanelle), in which the ultrasound
beam passes through soft tissue and a focal region is in the immediate vicinity of
bone. (unit-less)
TIC the CRANIAL BONE THERMAL INDEX.
TI
Sscan
the SOFT TISSUE THERMAL INDEX in a scanning mode. (unit-less)
TIS
non-scan
the SOFT TISSUE THERMAL INDEX in a non-auto scanning mode. (unit-less)
td the PULSE DURATION associated with the TRANSMIT PATTERN giving rise to the
reported value of MI. (microseconds)
z_at_max_Ipi,
α
the axial distance from the transducer where the ATTENUATED PULSE INTENSITY
INTEGRAL Ipi,α is maximum.
z
b
the distance where TIB_non-scan is determined. For non-scanned modes, Dis-
tance along the beam axis to the plane where the product of the ATTENUATED
OUTPUT POWER and ATTENUATED TEMPORAL-AVERAGE INTENSITY (Pα (z) x Ita,α
(z)) maximizes. (centimeters)
z
bp
value equal to 1.5 times the EQUIVALENT APERTURE DIAMETER (Deq). Also equals
1.69 * (centimeters)
z
s
the distance where TIS_non-scan is determined. The axial distance corresponding
to the location of max[min (Pα (z), Ita,α (z)x1 cm2)], where z >= zbp. (centimeters)
Explanatory Notes
(a) This index is not required to this operating mode.
(b) This probe is not intended for adult transcranial uses.
(c) This formulation for TIS is less than that for an alternate formulation in this mode.
(d) The maximum index value is less than 1.0
* If the MI comes from a scanning mode Transmit Pattern (pulse), the ‘prr’ listed is the average
per second for the ‘worst case’ scan line. ‘prr’ for scanning modes is the product of the frame
rate and the number of pulse per line of the Transmit Pattern.
** The max TIS unscanned value is an ‘at_surface’ value and occurs for aperture > 1.0 cm
2
, OR
The max TIS unscanned value is a ‘below_surface’ value and occurs for aperture <= 1.0 cm
2
.
+ The max TIB for this combinational mode = the at_surface TIS_scanned value. The non-
scanned_TIB value indicated is the below_surface max. A “+” is used when the TIB value is
less than the TIS_scanned value for the operating condition.