Reference Manual 127
EVN4-9: Pulsed Doppler & 2D+Pulsed Doppler
Index Label
M.I.
TIS TIB
TIC
scan
non-scan
non-
scan
A
aprt
≤1 A
aprt
>1
Maximum Index Value 1.400 0.594 0.617 0.483** 1.420 (b)
Associated
Acoustic
Parameters
P
r, α
(MPa)
2.940 – – – – –
P
(mW)
–
2D P1:2.190
PD P1x1:21.920
PD P1x1:
25.664
–
PD P:
19.300
(b)
Min.of
[P
α
(zs), Ita,
α
(z
s
)x1cm
2
]
(mW)
– – – 20.105 – –
z
s
(cm)
– – – 0.700 – –
z
bp
(cm)
– – – 1.043 – –
z
b
(cm)
– – – – 0.700 –
z_at_max_Ipi, α
(cm)
0.600 – – – – –
d
eq
(z
b
)
(cm)
– – – – 0.239 –
f
awf
(MHz)
4.420
2D :5.620
PD :5.130
5.050 5.050 5.160 (b)
Dim of A
aprt
X (cm)
–
2D :1.730
PD :0.633
0.633 0.633 0.295 (b)
Y (cm)
– 0.600 0.600 0.600 0.600 (b)
Other
Information
t
d
(µsec)
0.829 – – – – –
prr
(pulses/sec)
999.000 – – – – –
P
r
at max. I
pi
(MPa)
3.020 – – – – –
d
eq
at max. I
pi
(cm)
– – – – 0.209 –
I
pa, α
at max. MI
(W/cm
2
)
309.000 – – – – –
Focal Length
FLx (cm)
– – 4.000 4.000 – –
FLy (cm)
– – 2.300 2.300 – –
Operating
Control
Conditions
Control 1 MI
Control 2 TIS_as
Control 3 TIS_as_U TIS_bs
Control 4 TIB_bs
Control 5 TIC_as
Control 1: PulsedDoppler Mode, PD:4.4MHz,Focus:2.0Cm,Data No:3109721336-563195
Control 2: 2D+PulsedDoppler Mode, 2D:6.2MHzPD:5.1MHz,Focus:4.0Cm,FR:20.300,Data No:3110253259-717378
Control 3: PulsedDoppler Mode, PD:5.1MHz,Focus:4.0Cm,Data No:3109750355-515051
Control 4: PulsedDoppler Mode, PD:5.1MHz,Focus:2.0Cm,Data No:3109757202-619709
Control 5: PulsedDoppler Mode, PD:5.1MHz,Focus:1.0Cm,Data No:3109756048-259922
FDA Tables
Symbols and Denitions
All table entries have been obtained at the same operating conditions that give rise to
the maximum Index Value in the second row. Due to the complexities of the system user
interface, it may be dicult to exactly replicate the declared condition. Contact Samsung
Medison for further information.
Symbols used in the table are described below.
MI the Mechanical Index.
TIS
scan
the Soft Tissue Thermal Index in an auto-scanning mode.
TIS
non-scan
the Soft Tissue Thermal Index in a non-auto-scanning mode.
TIB the Bone Thermal Index.
TIC the Cranial Thermal Index.
A
aprt
the area of the active aperture (square centimeters).
p
r.3
the derated peak rarefactional pressure associated with the transmit pattern giving
rise to the value reported under MI (meg zapascals).
W
o
For TIB and TIC: time average acoustic power at the source, in milliwatts. (Also see
the denitions for W
01
and W
01x1
that follow.)
For TIS scan, W
o
= W
o1
+ W
o1x1
For TIS non–scan, W
o
= W
o1x1
W
o1
:
For scanning modes and/or scanning components of combinational modes:
time average acoustic power at the source, per cm, in milliwatts. This is the
acoustic power emitted from the central 1–cm length, in the scan direction,
of the aperture corresponding to the scanned pulses.