1-13
Chapter 1 Safety
Immunity test
IEC 60601
Test level
Compliance
level
Electromagnetic
environment - guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz
to 80MHz
0.01V Portable and mobile RF communications equipment
should be used no closer to any part of the
Ultrasound System, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
80MHz to 800MHZ
800MHz to 2.5GHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz
to 2.5GHz
3V/m where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recomm-ended separation
distance in meters (m).
Field strengths from xed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol :
NOTE 1) At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and
reection from structures, objects and people.
a
Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to xed RF transmitters, an electromagnetic site survey
should be considered. If the measured eld strength in the location in which the Ultrasound System is used exceeds
the applicable RF compliance level above, the Ultrasound System should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the
Ultrasound System or using a shielded location with a higher RF shielding eectiveness and lter attenuation.
b
Over the frequency range 150kHz to 80MHz, eld strengths should be less than [V
1
] V/m.