9-18
Operation Manual
Using the Biopsy Kit
WARNING:
X
Verify the condition of the biopsy needle before use. Do not use a bent biopsy needle.
X
The biopsy needle may bend during tissue penetration. The precise location of the needle
must be checked by monitoring the echo generated from the needle.
X
Never use the biopsy kit to biopsy prostate tissue.
Before Using Biopsy Kit
Ultrasonographic scanning using the biopsy kit must be conducted by medical doctors or experienced
medical sta with appropriate qualications. Always, without fail, verify all safety prevention
procedures and disinfection.
Use only Samsung Medison-approved biopsy kits. Other bands may not properly t Samsung Medison
probes. Improper installation may result in patient discomfort.
Inspect all components. Ensure that the biopsy kit you are using is the correct one for the probe, the
system, and system software.
WARNING:
X
Do not attempt to use the biopsy until you read the instructions for installing the sheath and
verifying alignment of the needle guide.
X
Always ensure that the probe and the needle guide are secured on both the left and the right.
X
Do not use in IVF, CVS, or PUBS procedures.
Biopsy Procedure
The system generates a needle guideline though the displayed real-time ultrasound images to indicate
the anticipated path of the needle. You can use this guideline to ensure that the needle or instrument
is following the correct path.
1. Ready the patient according to the procedure appropriate for the examination objectives.
2. Install the sheath and the biopsy kit.
3. Set the system controls for the biopsy procedure.If necessary, apply acoustic gel to the patient.
4. Begin scanning the patient. Adjust the patient so that the location for examination ts into the
needle guideline on the screen.