4-19
Chapter 4 Diagnosis Modes
X
Tag Pos: It is available when Color Tag is on. Use the Menu dial-button to select the position of
tags, from between 0 - 248. Tag Pos aect to Tag Width.
X
Tag Width: It is available when Color Tag is on. Use the Menu dial-button to select the range of
tags, from between 8 - 256. Tag Width aect to Tag Pos.
NOTE: For information on other menu items, please refer to “2D Mode.”
C Mode Softmenu
Following items are displayed on Softmenu in C mode. Select the item and adjust the value with the
Softmenu switch [1] – [5] on the control panel.
■
Scale
Use the Softmenu switch [1]. Lifting the switch makes the PRF (Pulse Repetition Frequency) increase
so that the speed range of the blood ow is widened, and vice versa.
■
Baseline
Use the Softmenu switch [2]. When the display method for Color Mode is set to Vel, the baseline
can be adjusted.
■
Filter
Use the Softmenu switch [3]. Filter is an electrical lter used to eliminate low-frequency Doppler
signals caused by the motion of vessel walls. Adjust Cuto Frequency to remove from the screen
the Doppler signals for which the frequencies are lower than the Cuto Frequency. Lift up the
switch to raise cuto frequency, and vice versa.
■
Steer
NOTE: This item appears in the Softmenu only when a Linear Probe is used.
Use the Softmenu switch [5]. Adjust the angle of the ultrasound beam. This can minimize loss of
color information based on the angle of the ultrasound beam. Press the switch to select Left, Right,
or None for ROI.
NOTE: For information on other Softmenu items, please refer to “2D Mode.”