8-6
Operation Manual
Fuse Replacement
The power protection fuse protects the product from excess current. If the power monitoring protection
circuit detects excess current, it shuts o the current to the equipment in order to prevent overheating
and to restrict the ultrasound power output.
If the fuse blows, replace it as shown below.
CAUTION: To avoid risk of electric shock, always disconnect the plug from the system prior to
fuse replacement.
AC Power Receptacle
Fuse
1
2
[Figure 8.1 Fuse replacement]
1. Turn o the system. (See u)
2. Take the fuse out by pressing the middle of fuse. (See v)
3. Insert the fuse holder into the appliance inlet.
4. After installing the new fuse, connect the power cord to the system.
Fuse information is shown in the following table.
Input Ratings Fuse Ratings Company Order No.
100-120VAC 10AH/250V Orisel 55T210000
200-240VAC 10A/250V Orisel 55T210000