5-20
Operation Manual
In case of measuring distance, choose a measurement method from Line or Trace. When
circumference and area are measured, choose a measurement method from Ellipse and Trace.
■
Rm Data
Press the Softmenu switch [2] Rm Data or the Clear button on the control panel. It erases all the
measurement data.
NOTE: If you use the Clear button, the measurement results are deleted from the screen but still
shown on the report for the corresponding application.
■
Cancel / Undo
Press the Softmenu switch [3] Cancel / Undo to cancel the latest measurement.
■
Rslt Disp Cnt
Lift up or down the Softmenu switch [4] Rslt Disp Cnt. Select number of the measurement result to
display. The information related to the item number is displayed in the screen for about a second.
Common Measurement Methods
This section provides information on the common measurement methods used for applications.
Measurements in Spectral Doppler Mode
In general, if you trace a Doppler spectrum, you can obtain results for various measurement items
automatically. There are 3 ways to trace a Doppler spectrum.
SONOACE R5 also allows you to select a specic item under the Measurement menu and take
measurements individually without tracing a Doppler spectrum.
■
Auto Trace
A spectrum is traced automatically. It is enabled in the Measurement menu in Spectral Doppler
Mode.
1. Press Auto Trace in the Measurement menu.
2. The system traces a spectrum automatically.
3. When Trace is complete, the measurement results are displayed on the screen.