9-18
Operation Manual
Disinfection
Disinfect the probe using a disinfectant solution recommended by Samsung Medison to reduce
pathogens to the level of 10
-6.
WARNING:
X
If a pre-mixed solution is used, be sure to observe the solution expiration date.
X
The type of tissue it will contact during use dictates the level of disinfection required for a probe.
Ensure that the solution strength and duration of contact are appropriate for disinfection.
CAUTION:
X
Using a non-recommended disinfectant or not following the recommended disinfection
method can damage and/or discolor the probe and will void the probe warranty.
X
Do not immerse probes for longer than one hour, unless they are sterilizable.
X
Only use liquid solutions to sterilize probes. Do not sterilize the probe using autoclave or EtO
gas.
1. Please refer to the user instructions of the disinfectant for details of proper storage, use, and disposal
of the disinfectant.
2. Mix the disinfectant compatible with your probe according to label instructions for solution
strength.
3. Immerse the probe into the disinfectant as shown in the illustration below.
4. Follow the instructions on the disinfectant to complete the immersion procedure of the probe, and
rinse the probe afterwards.
5. Allow the probe to air dry or wipe it with a dry cloth.
[Figure 9.1 Disinfecting a Probe]