9-20
Operation Manual
Using the Biopsy Kit
WARNING:
X
Verify the condition of the biopsy needle before use. Do not use a bent or warped biopsy
needle.
X
The biopsy needle may bend during tissue penetration. The precise location of the needle must
be checked by monitoring the echo generated from the needle.
X
Never use the biopsy kit to biopsy prostate tissue.
Before Using Biopsy Kit
WARNING:
X
Do not attempt to use the biopsy kit until you have read the instructions for installing the sheath
and verifying alignment of the needle guide.
X
Always ensure that the probe and the needle guide are secured on both the left and the right.
X
Do not use in IVF, CVS, or PUBS procedures.
Ultrasonographic studies using a biopsy kit must be conducted by a physician or other qualied
healthcare professional. Always, without fail, verify all safety procedures and disinfection.
Biopsy kits not supplied by Samsung Medison may fail to t the Samsung Medison probe properly.
Always use biopsy kits supplied by Samsung Medison, since improper setup may adversely aect the
patient.
Inspect all components of the biopsy kit. Ensure that the biopsy kit you are using is the correct one for
the probe, the system, and the system software.