SAMSUNG MEDISON DIAGNOSTIC ULTRASOUND SYSTEMUGEO PT60AUser Manual

Página 2

Precautions For UseBe sure to read this manual thoroughly to familiarize yourself with the operation of the product and the relevant safety informat

Página 3

4-4Operation ManualBasic Operating InstructionsThe items that are commonly used in each diagnosis mode are shown below:  Using the Touch Screen Menus

Página 4

4-5Chapter 4 Diagnosis Modes Q ScanNOTE: The Quick Scan function is only available with specific probes and applications.Use the Q Scan button on t

Página 5

4-6Operation Manual Basic Mode2D ModeThis basic mode, also referred to as B Mode (Brightness mode), provides scan planes of organs. This is used to d

Página 6

4-7Chapter 4 Diagnosis Modes[Figure 4.2 2D Mode - More]  Entering 2D ModeNOTE: Because 2D Mode is applied by default for all diagnosis modes, it can

Página 7

4-8Operation Manual Freq.(Frequency)This enables you to congure the probe’s frequency. Tap Freq. on the touch screen. You may select from Res, Pen,

Página 8 - Safety Requirements

4-9Chapter 4 Diagnosis Modes ClearVisionNOTE: ClearVision is an optional feature of this product.You can tap ClearVision on the touch screen to sel

Página 9 - Declarations

4-10Operation Manual MultiVisionNOTE: MultiVision is only displayed in the touchscreen menu when you are using a linear probe.Tap MultiVision on the

Página 10 - Precautions For Use

4-11Chapter 4 Diagnosis ModesBiopsy GuideSelect Biopsy on the touch screen.NOTE: Make sure to adjust the biopsy guideline before using the biopsy fea

Página 11

4-12Operation Manual  Starting and Finishing a BiopsyNOTE: Make sure to adjust the biopsy guideline before using the biopsy feature.1. Press the Bio

Página 12 - Revision History

4-13Chapter 4 Diagnosis ModesM ModeThe M Mode is used to specify an observation area in a 2D image with the M Line, and display changes over time.Thi

Página 13 - Table of Contents

DANGER: Disregarding this instruction may result in death, serious injury, or other dangerous situations.WARNING: Follow these instructions to prevent

Página 14 - Chapter 3 Starting Diagnosis

4-14Operation Manual  M Mode Menu Sweep SpeedUse the Sweep Speed ‘-/+’ buttons to select either 20mm/s, 30mm/s, 40mm/s, or 60mm/s. Time Line Only

Página 15 - Chapter 6 Image Management

4-15Chapter 4 Diagnosis ModesColor Doppler ModeThis mode displays the colored blood ow pattern of the ROI (Region of Interest) within the 2D image.I

Página 16 - Chapter 8 Maintenance

4-16Operation Manual  C Mode Screen ROI BoxROI stands for Region of Interest. The ROI Box outlines the area of the 2D image where color (blood ow)

Página 17 - Chapter 9 Probes

4-17Chapter 4 Diagnosis Modes ScaleTap the Scale button on the touch screen to use. Tapping ‘+’ of the Scale button makes the PRF (Pulse Repetition

Página 18

4-18Operation Manual[Figure 4.7 Color Doppler Mode - More]

Página 19 - Chapter 1

4-19Chapter 4 Diagnosis ModesPower Doppler ModeThis mode provides a color representation of blood ow intensity within the ROI in a 2D image. It is s

Página 20

4-20Operation Manual ROI BoxThe ROI (Region of Interest) outlines the area of the 2D image where color (blood ow) information is displayed in Power

Página 21 - Indication for Use

4-21Chapter 4 Diagnosis ModesPW Spectral Doppler Mode PW stands for Pulse Wave. This mode displays the blood ow speed at a specic blood vessel loca

Página 22 - Safety Information

4-22Operation Manual[Figure 4.11 PW Spectral Doppler Mode - More]  Entering & Exiting PW Spectral Doppler ModeTap the PW button on the touch scre

Página 23 - Chapter 1 Safety

4-23Chapter 4 Diagnosis Modes Adjusting Doppler BaselineTap the Baseline ‘+/-‘ buttons on the touch screen to adjust the Doppler baseline.  PW Spe

Página 24

Revision HistoryThe revision history of this operation manual is as follows:VERSION DATE NOTEv1.01.00-00 2014.06.18 Initial ReleaseProduct Upgrade a

Página 25 - Electrical Safety

4-24Operation Manual Combined ModeIn Combined Mode, three dierent modes are combined, including the default 2D Mode. Note that, in 2D/C Live Mode, o

Página 26

4-25Chapter 4 Diagnosis Modes Multi-Image ModeProvides Dual mode.In Multi-Image Mode, an image can be displayed in dierent combined modes. Button o

Página 28

Measurement Accuracy ...5-3Causes of Measurement Errors ...5-3Optimization of

Página 30

5-3Chapter 5 Measurements and Calculations Measurement AccuracyMeasurement values can vary depending on the nature of the ultrasound wave, the body’

Página 31

5-4Operation Manual  Ultrasound VelocityXThe average velocity of ultrasound used for measurement is usually 1,540m/s.XThe velocity of ultrasound ma

Página 32

5-5Chapter 5 Measurements and CalculationsOptimization of Measurement Accuracy  2D ModeXResolution is in proportion to the frequency of the probe.

Página 33

5-6Operation Manual  Color/Power Doppler ModeXA protocol is not specied for images in Color Doppler Mode or Power Doppler Mode. Therefore, the same

Página 34

5-7Chapter 5 Measurements and CalculationsMeasurement Accuracy TableThe following tables show the accuracy of the measurements available using the pr

Página 35 - Mechanical Safety

13Table of ContentsTable of ContentsChapter 1 SafetyIndication for Use ...

Página 36 - Moving the Equipment

5-8Operation Manual Basic MeasurementsTap the Meas. button on the touch screen’s soft menu.NOTE: Take basic measurements of distance and area regardl

Página 37 - Precautions for Use

5-9Chapter 5 Measurements and Calculations  Basic Measurement OperationsThe following is the information on common button operations for basic measu

Página 38

5-10Operation Manual[Figure 5.1 Caliper]Distance measurement  DistanceThis is a basic measurement that is available in all diagnosis modes. You can s

Página 39 - Biological Safety

5-11Chapter 5 Measurements and CalculationsTips!Repositioning PointYou may drag and drop the start point or the end point to change their position af

Página 40 -  Direct Controls

5-12Operation Manual  Distance (M)This is a basic measurement that is only available in M Mode. You can specify two points in an M image and measure

Página 41 -  Indirect Controls

5-13Chapter 5 Measurements and Calculations  VelocityThis is a basic measurement that is only available in Spectral Doppler Mode. You can specify on

Página 42 -  Output Display Features

5-14Operation ManualXPS/ED: PS ED RatioXED/PS: ED PS RatioThe formulae used for Manual Trace measurement are as follows:XDuration of flowVTImeanV =

Página 43 - Thermal Index (TI) Display

5-15Chapter 5 Measurements and CalculationsThe equations used for D Limited Trace measurement are as follows:XDuration of flowVTImeanV =XPSEDPSRI−=

Página 44

5-16Operation Manual  HRThis is a basic measurement that is only available in Spectral Doppler Mode. You can specify two bars on a Spectral Doppler i

Página 45 -  Color and Power Controls

5-17Chapter 5 Measurements and CalculationsThe following equation is used for Ellipse measurement: , (a: Half Long axis, b: Half Short axis)Area =, (

Página 46 -  Other

14Operation ManualBattery Pack ...

Página 47 -  Related Guidance Documents

5-18Operation ManualVolume measurementNOTE: Since Dual Mode simultaneously displays two images on the screen, you don’t have to return to the diagnosi

Página 48

5-19Chapter 5 Measurements and Calculations Calculations by ApplicationTap the Meas. button on the touch screen’s soft menu, and tap the Calc. butto

Página 49 - -(0,069 lf)

5-20Operation Manual Delete Trace LineDelete a trace line by reversing the dragging direction. You can only use this function when you are Manually

Página 50

5-21Chapter 5 Measurements and CalculationsCommon Measurement MethodsThis section provides information on the common measurement methods used for app

Página 51

5-22Operation Manual1. Select Auto Trace from the measurement menu. The system will automatically trace the spectrum.2. Two bars will appear at the

Página 52 - Systematic Uncertainties

5-23Chapter 5 Measurements and CalculationsOB CalculationsNOTE: XFetal HR can only be measured in Doppler Mode.XFor information on basic measuremen

Página 53 -  Training

5-24Operation ManualMeasurement Menu Item Mode Method UnitFetal BiometryGS 2D Distance measurement cmCRL 2D Distance measurement cmBPD 2D Distance m

Página 54 - Protecting the Environment

5-25Chapter 5 Measurements and Calculations  Calculating Estimated Fetal Weight (EFW)When measurements for the following items are complete, the sys

Página 55 - Battery Pack

5-26Operation ManualGynecology Calculations NOTE: XFor information on basic measurement methods, refer to ‘Basic Measurements’ and ‘Common Measuremen

Página 56

5-27Chapter 5 Measurements and CalculationsMeasurement Menu Item Mode Method UnitFollicle#1 All Volume measurement ml#2 All Volume measurement ml#3 A

Página 57 - Introduction

15Table of ContentsChapter 5 Measurements and CalculationsMeasurement Accuracy ...

Página 58

5-28Operation ManualVascular CalculationsThe measurement method for each item is the same as for basic measurement. NOTE: XFor information on basic m

Página 59 - Product Specications

5-29Chapter 5 Measurements and CalculationsAuto IMT Measurement1. After checking the probe and the application, start a carotid scan.2. If the desi

Página 60

5-30Operation ManualEchocardiographic Measurement (Cardiac Calculations)The measurement method for each item is the same as for basic measurement. Mea

Página 61 - Product Conguration

5-31Chapter 5 Measurements and Calculations ReportThe measurement results are summarized by application, and shown on the screen in a report format.V

Página 62 -  Touch Screen

5-32Operation Manual  Measurement Display MethodThe product allows you to measure one measurement item several times. However, only the rst three me

Página 63 - Chapter 2 Introduction

5-33Chapter 5 Measurements and CalculationsAdding CommentsTap Comment on the touch screen’s soft menu to switch to the screen for entering opinions o

Página 64 -  Touch Panel

5-34Operation ManualPrinting ReportsTap the Print button on the soft menu. The measurement report for the selected application will be printed on the

Página 65 - Console

Cine / Loop ...6-3Annotating Images ...6-5Annotation .

Página 67 - Peripheral Devices

6-3Chapter 6 Image Management Cine / LoopImages are automatically saved in the memory during scanning. The saved images are useful for patient diagn

Página 68 - WLAN Card

16Operation ManualChapter 7 UtilitiesSystem Settings ...

Página 69 - Video Out

6-4Operation Manual  How to Review ImagesMove the cursor to the bar to review an image. You can search through saved images in chronological order by

Página 70

6-5Chapter 6 Image Management Annotating ImagesAnnotationEnter or display text, Arrows, or BodyMarkers on an image. This function is useful for die

Página 71 - Battery Pack

6-6Operation Manual  Entering Text 1. Press the Annotation button on the soft menu, and then press the Text button to switch to Text Entry mode.2.

Página 72 -  Battery Icons

6-7Chapter 6 Image Management SpaceEnter a space in the Text Entry window. BackSpaceDelete the previous character. ExitExit the Text Library me

Página 73 -  Connecting Probes

6-8Operation Manual  Entering Arrow 1. Press Annotation button on the soft menu, and then press the Arrow button to switch to entry mode.2. To adju

Página 74 - UGEO PT60A CART (Optional)

6-9Chapter 6 Image ManagementBodyMarkerEnter a BodyMarker on an image. This function is useful for dierentiating or marking a diagnosis area.  Star

Página 75 - Accessories

6-10Operation Manual  Changing the Probe Cursor AnglePress the Rotate button on the soft menu.  Repositioning BodyMarker1. Press the Move Marker bu

Página 76

6-11Chapter 6 Image Management Saving, Playing and Transferring ImagesSaving ImagesWARNING: You must enter the patient ID because all images are sav

Página 77 - Optional Functions

6-12Operation ManualTransferring ImagesUGEO PT60A allows you to transfer images to a PACS system that supports DICOM. You can transfer all saved image

Página 78

6-13Chapter 6 Image Management SonoView ModeSonoView is UGEO PT60A's integrated image management program. It oers image saving/ling, preview,

Página 79 - Starting Diagnosis

17Table of ContentsChapter 9 ProbesProbes ...

Página 80

6-14Operation Manual  Launching SonoView ModePress the SonoView button on the soft menu. The screen will switch to the SonoView screen.If there are

Página 81 - System Power

6-15Chapter 6 Image ManagementExam ModePress the SonoView button on the soft menu. In the Exam Mode, you can review the current or saved exam.[Figure

Página 82 - Shutting Down the System

6-16Operation Manual  Selecting ImageSelect images in the Thumbnail List. The selected image is highlighted in yellow in the list. Specify the positi

Página 83 - Probes and Presets

6-17Chapter 6 Image Management  Transferring Images via DICOMTransfer the selected image and measurement data via DICOM. The icon is only enabled wh

Página 84 - Managing Presets

6-18Operation Manual  Deleting ImagesPress the Delete button to delete the selected image.  Exiting SonoViewPress the Exit button on the soft menu.

Página 85 -  Overwriting Presets

System Settings ...7-3General System Settings ...7-

Página 87

7-3Chapter 7 Utilities System SettingsGeneral system settings that do not aect the image are described. These settings may be changed as needed o

Página 88

7-4Operation ManualGeneral System SettingsOn the Setup screen, tap the System button on the touch screen. You can specify general settings such as tit

Página 89

7-5Chapter 7 Utilities  Date and TimeCongure the system's date and time s ettings.Today's date is displayed. To change the date, tap .N

Página 91 -  Pediatric

7-6Operation ManualScreen Display Settings (Imaging)On the Setup screen, tap the Imaging button on the touch screen to congure settings for image dis

Página 92 -  Cardiac

7-7Chapter 7 Utilities  Preset Settings Thermal IndexSpecify the TI to display on the screen as TIs (Soft tissue Thermal Index), TIb (Bone Therm

Página 93 -  Emergency

7-8Operation Manual  Arrow Arrow SizeSelect the arrow size (Small or Large). Arrow TypeSelect the arrow shape (Arrow or Hand).Measurement Setting

Página 94 - Searching Patient Information

7-9Chapter 7 Utilities  Heart Rate CycleSet the number of beats for the heart rate (HR) cycle. This is the setting for the HR Cycle that will be

Página 95 -  Worklist Search

7-10Operation Manual Gestational AgeSpecify the Fetal Age Author for each Label. Fetal GrowthSpecify the Fetal Growth Author for each Label.  EFW

Página 96

7-11Chapter 7 Utilities Peripheral Device SettingsOn the Setup screen, select the Periph. button to congure the peripheral devices connected to the

Página 97 - Diagnosis Modes

7-12Operation Manual  Store StoreStores the image. Store/printStores and prints the image.  Brightness Screen BrightnessSelect the screen brig

Página 98

7-13Chapter 7 Utilities Network SettingsOn the Setup screen, tap the Net button to congure the general settings for wired and/or wireless networks.

Página 99 - Information

7-14Operation Manual  WLANUse a USB adapter to connect the system to a wireless network. NOTE: The WLAN Settings window is enabled only when the syst

Página 100 - Basic Operating Instructions

7-15Chapter 7 Utilities DICOM Settings Select the DICOM button on the Setup screen.NOTE: For more information, please refer to the server’s user man

Página 101 - Chapter 4 Diagnosis Modes

Indication for Use ...1-3Contraindications ...

Página 102 - Basic Mode

7-16Operation Manual AE TitleEnter a name for the DICOM AE (Application Entity). This is used to identify devices that use DICOM on the network. P

Página 103 -  2D Mode Menu

7-17Chapter 7 Utilities  Port No.Enter the Port Number for the server being used. Please contact your organization’s network administrator to nd

Página 104 - Frame Average

7-18Operation ManualStorage Server InformationSelect the ‘STORAGE’ item from the combo boxes at the top right corner. Set up the image storage service

Página 105

7-19Chapter 7 Utilities Storage SR Server InformationSelect the ‘STORAGE SR’ item from the combo boxes at the top right corner. Set up the DICOM ser

Página 106 - MultiVision

7-20Operation ManualServiceOn the Setup screen, tap the Service button. Users should not attempt to perform anything on their own. They must contact t

Página 107 - Biopsy Guide

7-21Chapter 7 Utilities InfoOn the Setup screen, tap the Info button; the software version information for the product will be displayed. You may vi

Página 109 -  M Mode Screen

Operational Environment ...8-3Product Maintenance ...8-4Cleaning and disinfecting .

Página 111 - Color Doppler Mode

8-3Chapter 8 Maintenance Operational EnvironmentWhen installing the product, please pay attention to the following information.CAUTION: If the pro

Página 114 - Operation Manual

8-4Operation Manual Product MaintenanceCleaning and disinfectingUsing an inappropriate cleaning or sterilizing agent may damage the product. Take the

Página 115 - Power Doppler Mode

8-5Chapter 8 Maintenance  DisinfectingCAUTION: When disinfecting the product's exterior, be sure to use disinfectants recommended by Samsung

Página 116 -  PD Mode Menu

8-6Operation Manual Battery Pack ManagementBattery packs are consumables, and their capacity to hold charge gradually decreases as they are used. Whe

Página 117 - PW Spectral Doppler Mode

8-7Chapter 8 Maintenance  Connecting a Battery PackConnect the battery pack as follows:1. Turn o the system and disconnect the power cord from t

Página 118

8-8Operation Manual [Figure 8.2 Docking to Cart]The battery pack must be charged and discharged within the following temperature ranges:State Ambient

Página 119 - Wall Filter

8-9Chapter 8 Maintenance Storing the Battery PackIf you are not planning on using your UGEO PT60A, be sure to remove the battery pack from the produ

Página 120 - Combined Mode

8-10Operation Manual Information MaintenanceCAUTION: Back up the data on the system on a regular basis, since an external shock or an internal error

Página 121 - Multi-Image Mode

Probes ...9-3Ultrasound Transmission Gel ...

Página 123 - Calculations

9-3Chapter 9 Probes ProbesThe probe is a device that sends and receives ultrasound for acquiring image data.It is also called a Transducer or Scanh

Página 124

1-3Chapter 1 Safety Indication for UseThe UGEO PT60A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and

Página 125 - Measurement Accuracy

9-4Operation ManualProbe Application PresetC2-8OB OBAbdomen AbdomenGynecology GynecologyC2-5OB OBAbdomen AbdomenGynecology GynecologyLN5-12Small Parts

Página 126 -  Human Error

9-5Chapter 9 Probes  Function ListThe functions you can use on UGEO PT60A for each probe and presets are as follows: Probes Application Q Scan Har

Página 127 -  Doppler Mode

9-6Operation ManualProbes Application Q Scan Har MultiVision ClearVision SDMR BiopsyPN2-4Cardiac O O X X O XEmergency O O X X O XAbdomen O O X X O XTC

Página 128 -  Cursor Position

9-7Chapter 9 Probes  Thermal Index (TI) TablesThe Thermal Index (TI) is displayed in the Title area of the screen, and represents the probability

Página 129 - Measurement Accuracy Table

9-8Operation ManualUltrasound Transmission GelFor successful acoustic signal transmission, please only use the ultrasound gels approved by Samsung Med

Página 130 - Basic Measurements

9-9Chapter 9 Probes Using SheathsSheaths are recommended for clinical applications of invasive nature, including intraoperative, transrectal, transv

Página 131 - Exiting Basic Measurement

9-10Operation ManualProbe Safety PrecautionsCAUTION: XDo not apply mechanical shock to the probe.XDo not place the probe cable on the floor where th

Página 132 - Distance measurement

9-11Chapter 9 Probes  Electric ShocksThe probe uses electrical energy. If it touches conductive materials, there are risks of electric shocks to t

Página 133 -  Angle

9-12Operation ManualCleaning and Disinfecting the ProbeWARNING: XAlways use protective eyewear and gloves when cleaning and disinfecting probes.XIns

Página 134 -  Slope

9-13Chapter 9 Probes NamesDisinfectantsT-Spray IIT-SpraySani-Cloth HBSani-Cloth PlusSani-Cloth ActiveSeptiwipesCleanisept WipesSter-Bac BluTransepti

Página 135 -  Manual Trace

1-4Operation Manual Safety InformationPlease read the following safety information before using this product. It is relevant to the ultrasound system

Página 136 -  Limited Trace

9-14Operation ManualNamesDisinfectantsCidex OPA 2,3,6)Cidex Plus 2)Metricide 2)Omnicide (28)Omnicide 14NSOmnicide - FG2NucleanWavicide-01 3)Sekusept E

Página 137 -  Auto Trace

9-15Chapter 9 Probes NamesDisinfectants CleanerVirkonSporoxSporox IIGigaseptGigasept AF 3)Gigasept FFHibitanePeraSafeEnzolAlkazymeCidezymeKlenzymeIs

Página 138 -  Ellipse

9-16Operation ManualNamesCleaner GelEthanol 75%MetrizymeMcKessonNatural ImageAquasonics 100 3)GE Ultrasound Contact GelClear ImageKendallScanWavelengt

Página 139 -  Trace

9-17Chapter 9 Probes Tips!SymbolsThe significance of each symbol is as follows: (1) Compatible but no EPA Registration (2) FDA 510 (k) Cleared (3)

Página 140 - Volume measurement

9-18Operation ManualThe following is information about the manufacturers (or Distributors) of detergents, disinfectants, and ultrasound gels.Product M

Página 141 - Calculations by Application

9-19Chapter 9 Probes  CleaningCleaning is an important procedure that is carried out before disinfecting the probe. The probe must be cleaned aft

Página 142

9-20Operation Manual  DisinfectionDisinfect the probe using a disinfectant solution recommended by Samsung Medison to reduce pathogens to the level o

Página 143 - Common Measurement Methods

9-21Chapter 9 Probes BiopsyA biopsy is an examination method that surgically extracts tissue from the patient for examination. The probe and the bi

Página 144 -  Heart Rate Measurement

9-22Operation ManualUsing the Biopsy KitWARNING: XOnly use needles that have been approved for use in your country. XVerify the condition of the bio

Página 145 - OB Calculations

9-23Chapter 9 Probes  Biopsy ProcedureThe system generates a needle guideline using the ultrasound image displayed in real-time to anticipate the

Página 146 -  AFI(Amniotic Fluid Index)

1-5Chapter 1 SafetySymbols Description Symbols DescriptionPower on Microphone PortPower o Probe portPower ON for part of the productIPX 1: Protected

Página 147

9-24Operation Manual  Needle Guide AlignmentWARNING: XDo not use the same needle used to verify alignment for an actual biopsy application. Always u

Página 148 - Gynecology Calculations

9-25Chapter 9 Probes Assembling the Biopsy Kit  Reusable adapter1. Mount the adapter on the probe.2. Fill the sheath with the ultrasound gel and

Página 149

9-26Operation Manual  Disposable adapter1. Fill the sheath with ultrasound gel, and cover the probe completely with the sheath.2. Mount the adapter

Página 150 - Vascular Calculations

9-27Chapter 9 Probes  Biopsy Kit SpecicationsProbeBiopsyModel NameComponentsAdapter MaterialReusable / DisposableNeedle Gauge Multi AngleC2-5 BP-

Página 151 - Auto IMT Measurement

9-28Operation ManualCleaning and Disinfecting the Biopsy KitWARNING: Always use protective eyewear and gloves when cleaning and disinfecting biopsy ki

Página 152

9-29Chapter 9 Probes  Cleaning and Disinfecting of Plastic Biopsy Kit Cleaning 1. After use, remove the biopsy kit from the probe.2. Disassem

Página 154 -  Measurement Display Method

1-6Operation ManualLABELWarning and caution labels that contain information and instructions concerning the protection of the product can be found on

Página 155 - Store SR

1-7Chapter 1 Safety Electrical SafetyThis equipment has been veried as a Class I device with Type BF applied parts.Prevention of Electric ShocksWAR

Página 156 - Closing Reports

1-8Operation ManualCAUTION: XThe system has been designed for 100-240VAC; you should select the input voltage of any connected printer and VCR. Prior

Página 157 - Image Management

1-9Chapter 1 SafetyECG-Related InformationWARNING: XThis device is not intended to provide a primary ECG monitoring function, and therefore does not

Página 158

1-10Operation ManualEMIAlthough this system has been manufactured in compliance with existing EMI (Electro-Magnetic Interference) requirements, use of

Página 159 - Cine / Loop

1-11Chapter 1 Safety  Approved Cables, Probes and Peripherals for EMC CablesCables connected to this product may aect its electromagnetic emissio

Página 161 - Annotating Images

1-12Operation ManualImmunity test IEC 60601 Test level Compliance levelElectromagnetic environment - guidanceElectrostatic discharge (ESD)IEC 61000-4-

Página 162 -  Text Library Menu

1-13Chapter 1 SafetyImmunity testIEC 60601test levelCompliance levelElectromagnetic environment - guidanceConducted RFIEC 61000-4-63 Vrms150 kHzto 80

Página 163 -  Arrow Mode

1-14Operation Manual  Recommended distance between wireless communication device and this productThis product is intended for use in an electromagnet

Página 164 -  Deleting Arrows

1-15Chapter 1 SafetyCAUTION: If the system is connected to customer-supplied equipment, such as a local area network (LAN) or a remote printer, Samsu

Página 165 - BodyMarker

1-16Operation ManualA medical device can either generate or receive electromagnetic interference. The EMC standards describe tests for both emitted an

Página 166 - Exiting Annotation

1-17Chapter 1 Safety Mechanical SafetyPrecautions for UseCAUTION: XDo not apply excessive force to the product. XNever attempt to modify the produ

Página 167 - Saving Images

1-18Operation ManualMoving the EquipmentFirmly grip the handle on the front of the product and move the product slowly. Alternatively, you can use the

Página 168 - Transferring Images

1-19Chapter 1 SafetyPrecautions for UseCAUTION: XDo not press the control panel excessively.XNever attempt to modify the product in any way.XCheck

Página 169 - SonoView Mode

1-20Operation Manual  Caution for Using Control PanelCAUTION: XDo not push the control panel with excessive force or lean on itXDo not sit on the c

Página 170 -  Launching SonoView Mode

1-21Chapter 1 Safety Biological SafetyFor safety information on the probe and biopsy, refer to ‘Chapter 9. Probes’.WARNING: XUltrasound waves may h

Página 172 -  Searching for an Exam

1-22Operation Manual  Applying ALARAThe system-imaging mode used depends upon the information needed. 2D-mode and M-mode imaging provide anatomical i

Página 173 -  Print

1-23Chapter 1 Safety  Indirect ControlsThe indirect controls are those that have an indirect eect on the acoustic intensity. These controls aect t

Página 174 -  Exiting SonoView

1-24Operation Manual  Additional ConsiderationsEnsure that scanning time is kept to a minimum, and ensure that only medically required scanning is pe

Página 175 - Utilities

1-25Chapter 1 Safety Mechanical Index (MI) DisplayMechanical bioeects are threshold phenomena that occur when a certain level of output is exceede

Página 176

1-26Operation ManualConservative values for tissue characteristics were selected for use in the TI models. Conservative values for tissue or bone abso

Página 177 - System Settings

1-27Chapter 1 SafetyIn combined modes, such as simultaneous Color, 2D-mode, and pulsed Doppler, the individual modes each add to the total TI. Each m

Página 178 - General System Settings

1-28Operation Manual ScaleUsing the scale control to increase the color velocity range may increase the TI. The system will automatically adjust th

Página 179 -  Screen Saver

1-29Chapter 1 Safety ProbesEach probe model available has unique specications for the contact area, beam shape, and center frequency. Settings are

Página 180 -  Common Settings

1-30Operation ManualXAcoustic Output Labeling Standard for Diagnostic Ultrasound Equipment (AIUM, 2008)XStandard Means for the Reporting of the Acou

Página 181 - Annotation Settings

1-31Chapter 1 Safety  In Situ, De-rated, and Water Value IntensitiesAll intensity parameters are measured in water. Since water does not absorb acou

Página 182 - Measurement Settings

MI68-02968ASAMSUNG MEDISON DIAGNOSTIC ULTRASOUND SYSTEMVersion 1.01EnglishUGEO PT60AUser Manual

Página 183 -  Author Settings

1-32Operation Manual  Terms and Symbols Related to Acoustic Output and MeasurementThe terms and symbols used in the acoustic output tables are dened

Página 184 -  EFW Settings

1-33Chapter 1 Safetyprr PULSE REPETITION RATE, the inverse number of the time interval between two contiguous acoustic pulse (unit: Hz)TIB BONE THERM

Página 185 - Peripheral Device Settings

1-34Operation Manual  Acoustic Measurement Precision and UncertaintyThe Acoustic Measurement Precision and Acoustic Measurement Uncertainty are descr

Página 186 -  Clip Store

1-35Chapter 1 SafetyXLinearity of the balance system.XExtrapolation to the moment of switching the ultrasonic transducer (compensation for ringing

Página 187 - Network Settings

1-36Operation Manual Protecting the EnvironmentCAUTION: XFor disposal of the system or accessories that have come to the end of their lifespan, cont

Página 188 - Password

1-37Chapter 1 Safety Battery PackFamiliarize yourself with the instructions below before using the battery pack:WARNING: XComply with all instructi

Página 189 - DICOM Settings

1-38Operation Manual Correct Disposal of Batteries in This ProductApplicable in countries with separate collection systemsThis marking on the battery,

Página 190 -  DICOM Service Settings

Product Specications...2-3Product Conguration ...2-5Monitor ...

Página 192 - PPS Server Information

2-3Chapter 2 Introduction Product SpecicationsPhysical DimensionsSYSTEM:Height: 252.6 mm, Width: 291 mm, Depth: 52.2 mmWeight: 3.6kg (Including bat

Página 194

2-4Operation ManualElectrical ParametersAC/DC adapter1) FSP180-ABAM1Input: 100-240VAC, 2.3-0.8A, 50/60Hz Output: 19VDC, 9.47A, 180VA 2) PMP220-13-2Inp

Página 195

2-5Chapter 2 Introduction Product CongurationThis product consists of a monitor, a console, peripheral devices, and probes.215431 Touch Panel2 Touc

Página 196

2-6Operation ManualMonitorThe monitor consists of a touch screen and a touch panel.  Touch ScreenFor this product, the touch screen displays ultrasou

Página 197 - Maintenance

2-7Chapter 2 Introduction4 Soft Menu AreaDisplays Soft Menu touch buttons that can be used in the current mode.Tips!Current System StatusIndicates th

Página 198

2-8Operation Manual  Touch PanelA control menu located in the glass part at the bottom of the monitor, with which you can set up patient information

Página 199 - Operational Environment

2-9Chapter 2 IntroductionConsole The console consists chiey of ultrasound imaging components on the inside, and various connectors and a handle on t

Página 200 - Product Maintenance

2-10Operation Manual  Console – Left SidePower Port  Console – Right SideProbe PortSecurity Lock Port (Kensington Lock)USB port  Console - Top11SD

Página 201 - Accuracy Checks

2-11Chapter 2 IntroductionPeripheral DevicesPeripheral devices can be connected to their corresponding ports on the left/right or rear sides of the c

Página 202 - Battery Pack Management

2-12Operation ManualThe following products are recommended: USB Video PrinterXColor: Mitsubishi CP30DW, SONY UP-D25MDXBlack and White: Mitsubishi

Página 203 - Recharging the Battery Pack

2-13Chapter 2 IntroductionVideo OutThe optional Video Out device connects to the system to provide digital and analog video outputs.The Video Out dev

Página 204

PROPRIETRAY INFORMATION AND SOFTWARE LICENSEThe Customer shall keep confidential all proprietary information furnished or disclosed to the Customer by

Página 205 - Disposing of the Battery Pack

2-14Operation ManualConnect the Video Out device to the console in the following order:1. Connect the AC adapter to the AC adapter connection port on

Página 206 - Information Maintenance

2-15Chapter 2 IntroductionBattery PackThe battery pack of this product contains lithium ion batteries. Use this battery pack if you are not using an

Página 207 - Chapter 9

2-16Operation Manual  Battery IconsIcons that indicate the battery status are displayed in the User Information Area on the screen. If you are using

Página 208

2-17Chapter 2 IntroductionProbesProbes are devices that generate ultrasound waves and process reected wave data for the purpose of image formation.N

Página 209 -  Probe List

2-18Operation ManualUGEO PT60A CART (Optional)NOTE: UGEO PT60A CART is an optional feature of this product.The UGEO PT60A CART is a convenient platfor

Página 210

2-19Chapter 2 IntroductionAccessoriesA box containing various accessories is supplied with this product.SONO GELSYSTEM POWER CABLESYSTEM ADAPTERCRADL

Página 211 - Function List

2-20Operation ManualCRADLEThe supplied cradle is manufactured only for UGEO PT60A. Please place the product on the cradle to perform an exam safely. [

Página 212

2-21Chapter 2 IntroductionOptional FunctionsThis product has the following optional features. XDICOMXClearVisionXAuto IMTFor further information a

Página 214 - Ultrasound Transmission Gel

System Power ...3-3AC Adapter Connection ... 3-

Página 215 - Using Sheaths

Safety Requirements Classifications: XType of protection against electrical shock: Class IXDegree of protection against electrical shock (Patien

Página 217 -  Electric Shocks

3-3Chapter 3 Starting Diagnosis System PowerBoot up the system for use.CAUTION: Before turning the power on, connect the probes and peripheral devic

Página 218

3-4Operation ManualTurning the Power On While the power is turned o, press the On/O button. Booting will start, and the product logo will be display

Página 219

3-5Chapter 3 Starting Diagnosis Probes and PresetsBefore starting a scan, select a probe and a preset.Tap the Probe button on the touch panel, and t

Página 220

3-6Operation ManualSelecting Probe and PresetSelect the items in the following order: Probe > Preset. Pressing Preset loads the settings and initia

Página 221

3-7Chapter 3 Starting Diagnosis  Renaming Presets Only user presets can be renamed.1. Tap Rename on the touch screen.2. Select the preset you wish

Página 222

3-8Operation Manual Patient Information Tap the Patient button on the touch panel, and the Patient screen will be displayed.In this screen, you can e

Página 223

3-9Chapter 3 Starting Diagnosis Patient IDEnter patient ID. Last NameEnter the patient’s last name. First Name Enter the patient’s rst name.

Página 224

3-10Operation Manual Auto IDCreate temporary patient information using a temporary ID. Start ExamStart an exam. WorklistPerform a search by conn

Página 225 -  Cleaning

3-11Chapter 3 Starting DiagnosisPatient Information for Each ApplicationIn Study Information, enter additional patient information or change the exis

Página 226 -  Disinfection

X Medical Electrical Equipment, Part 2-37: Particular Requirements for Safety - Ultrasonic Medical Diagnostic and Monitoring Equipment [IEC 60601-2-3

Página 227 - Biopsy Kit Components

3-12Operation Manual  GynecologyFrom Application, select Gynecology to enter additional information relevant to gynecology. The information entered i

Página 228 - Using the Biopsy Kit

3-13Chapter 3 Starting Diagnosis  AbdomenSelect Abdomen in Application. HeightEnter the patient’s height in centimeters (cm), feet (ft), inch (in)

Página 229 -  Biopsy Procedure

3-14Operation Manual  Small PartsSelect Small Parts in Application. HeightEnter the patient’s height in centimeters (cm), feet (ft), inch (in). W

Página 230 -  Needle Guide Alignment

3-15Chapter 3 Starting Diagnosis  AnesthesiaSelect Anesthesia in Application. HeightEnter the patient’s height in centimeters (cm), feet (ft), inc

Página 231 - Assembling the Biopsy Kit

3-16Operation ManualSearching Patient Information  SearchSearch through the information stored in the system.[Figure 3.4 Search]1. Select the Search

Página 232 -  Disposable adapter

3-17Chapter 3 Starting Diagnosis  Worklist SearchPerform a search by connecting to the DICOM Modality Worklist server in the hospital network.NOTE:

Página 234 - Disinfection

Information ...4-3Types of Diagnosis Modes ...4-3

Página 236

4-3Chapter 4 Diagnosis Modes InformationTypes of Diagnosis ModesIn addition to Basic Mode, Combined Mode and Multi-Image Mode are provided. Basic M