Vista de página 30
1-13
Chapter 1 Safety
Cable Type Length
Foot Switch Shielded 2.5m
B/W Printer Unshielded Coaxial Normal
MIC Unshielded Any
Printer Remote Unshielded Any
Audio R.L Shielded Normal
VHS Shielded Normal
ECG AUX input Shielded < 3m
Parallel Shielded Normal
Approved Transducer for Electromagnetic Compliance
The probe listed in ‘Chapter 9. Probes’ when used with this product, have been tested to comply
with the group1 class B emission as required by International Standard CISPR 11.
Approved Accessories for Electromagnetic Compliance
Accessories used with this product may effect its emissions.
CAUTION: When connecting other customer-supplied accessories to the system, such as a remote
printer or VCR, it is the users responsibility to ensure the electromagnetic compatibility of the
system. Use only CISPR 11 or CISPR 22, CLASS B compliant devices.
WARNING: The use of cables, transducers, and accessories other than those specified may result in
increased emission or decreased Immunity of the Ultrasound System.