1-12
Operation Manual
EMC
The testing for EMC(Electromagnetic Compatibility) of this system has been performed according to the
international standard for EMC with medical devices (IEC60601-1-2). This IEC standard was adopted in
Europe as the European norm (EN60601-1-2).
Guidance and manufacturer’s declaration - electromagnetic emission
This product is intended for use in the electromagnetic environment specified below. The customer or
the user of this product should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment -guidance
RF Emission
CISPR 11
Group 1
The Ultrasound System uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
RF Emission
CISPR 11
Class B
The Ultrasound System is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies building used for domestic purpose.
Harmonic Emission
IEC 61000-3-2
Class A
Flicker Emission
IEC 61000-3-3
Complies
Approved Cables, Transducers and Accessories for EMC
■
Approved Cable for Electromagnetic Compliance
Cables connected to this product may affect its emissions;
Use only the cable types and lengths listed below table.
Cable Type Length
VGA Shielded Normal
RS232C Shielded Normal
USB Shielded Normal
LAN(RJ45) Twisted pair Any
S-Video Shielded Normal