4-25
Chapter 4 Diagnosis Modes
■
Steer
NOTE: The Steer function appears on the soft menu only when a linear probe is used.
Adjust the angle of the ultrasound beam. Loss of color information resulting from the angle of the
ultrasound beam is minimized. Select Left or Right on the touch screen. Pressing the button again
cancels the selection.
■
Display Mode
Select C mode to use. Select Color + B/W or B/W Only by tapping the button on the screen or using
the dial-button.
■
Color Mode
Sets details of the color display. Select Vel (Velocity) or Vel+Var (Variance) by tapping the button on
the touch screen or using the dial-button. This button is enabled only in Phased Array Probe.
■
Balance
Select a value between 1 and 16 by tapping Balance on the touch screen or using the dial-button.
The range of a color image is adjusted by comparing the gray levels of 2D images with the doppler
signal values of color images. When the Balance value increases, the color image also appears in the
part where the gray level of a 2D image is high (the bright part), increasing the range of the color
image.
■
Color Map
Used to configure the post curve in a color image. You can visually highlight blood flow. Select a
value between 1 and 15 by tapping Color Map on the touch screen or using the dial-button.
■
Sensitivity
Select a value between 8 and 31 by tapping Sensitivity on the touch screen or using the dial-
button.
When this value increases, the sensitivity of a color image improves but the frame rate becomes
lower.