5-31
Chapter 5 Measurements and Calculations
OB Calculations
NOTE:
X
Ductus Venosus, Fetal HR can only be measured in Doppler Mode.
X
For information on basic measurements, please refer to “Basic Measurements” in this chapter.
X
For references on measurement items, please refer to “Reference Manual – Part 1.
Before Taking OB Measurements
OB Basic Information
Enter the information required for OB diagnosis in the Patient Information window. The basic OB
information includes LMP (Last Menstrual Period) and Gestations.
Once LMP is entered, EDD (Estimated Delivery Date) and GA (Gestational Age) are calculated
automatically. LMP is required for the calculation of values such as EDD and SD in obstetrics
measurement.
X
EDD(LMP) = LMP + 280 days
X
GA(LMP) = Current System Date - LMP
When you input Ovulation Date, it creates LMP and GA(LMP), EDD(LMP) automatically.
LMP=Ovul.Date-14
Regardless of LMP, enter the EDD with a physicians opinion into Estab. Due Date. If LMP is not
available, when Estab. Due Date is modified, LMP is automatically calculated and the “C” mark is
displayed next to the LMP information.
A maximum of four fetuses can be entered in the Gestations menu. The default value is ‘1’. In the
case of twins, enter ‘2’.
For further information about patient information menus and how to input patient information,
refer to “Entering Patient Data” in Chapter 3 “Setting”.
OB Measurement Menu Settings
Set up the GA Equation, GA Table and OB measurement menus that are used in obstetrics
measurements. The user can manually write, back up or restore GA Tables. For more information
on the GA Equation and Table, refer to the Reference Manual.