4-100
Operation Manual
Trimester
Set the pregnancy trimester. Select from 1
st
, 2
nd
and 3
rd
by clicking the corresponding button or using
the dial-button on the touch screen.
Tips!
Trimester
If 1st - 3rd are set, the recommended scan time and STIC angle are automatically set for the
specified trimester. Please see the following table:
Trimester
1st 2nd 3rd
ScanTime
10 seconds 12 seconds 15 seconds
STIC Angle
20˚ 25˚ 30˚
If a scan time and STIC angle other than the ones in the above table are set, the trimester is
displayed as user set.
NOTE: For information on other usage, please refer to “3D Stand By.
Reviewing XI STIC Image
XI STIC images are played as Volume Cine. Under XI STIC images, XI STIC information, including Img
Angle, Vol. Angle, Scan Time, HR, Volume Pos., are shown.
NOTE: The only available modes with Volume Cine are MPR, MSV and Oblique View.
Press the Freeze button on the control panel to stop Volume CINE playback.
Speed (%)
NOTE: This option is available only in MPR or MSV Mode during Volume Cine playback.
Set the playback speed for XI STIC images. Select between 25 and 400% by clicking the corresponding
button or using the dial-button on the touch screen. This rate is based on the fetal heart rate (100%).