Chapter 4 Diagnosis Modes
This function allows you to view more accurate images of organ or tissue boundaries. A higher
value provides more accurate images of boundaries.
■
Power
Tap Power on the touch screen. Select ultrasound power between 10 – 100 by tapping a desired
button or using the dial-button on the touch screen.
■
Tissue
Select from Solid, Normal, Adipose or Cystic by tapping a desired on the touch screen.
Low MI
NOTE: Low MI function will be displayed in the 2D menu under following conditions;
X
Probe: C1-4EC
X
Application: Contrast
X
Preset: Perfusion
This function is used to observe tissues with a contrast agent injected.
Tap Low MI in the touch screen to switch to the Low MI screen. The Low MI symbol appears in the
upper right corner of the screen and the Low MI timer appears in the upper left corner of the screen.
Tips!
Entering Low MI Mode
X
FSI turns off and Pulse Inversion turns on.
X
Other functions, such as Harmonic and PPI, become disabled.
X
You cannot switch to other diagnosis modes.
■
Low MI Menu
X
MI Control: Select the MI value from 1 to 60.
X
Flash: The image will be displayed brightly for the period of time set in Flash Time.
X
Flash Time: Set a value between 1 and 10 to specify the period of time for which the Flash
function will operate.
X
Start / Stop: Start or stop the Low MI timer.
X
Reset: Reset the Low MI timer.
X
Chroma: Tap the button to select On or Off. When turned on, the colors of the image displayed
on the screen will be changed arbitrarily. Chroma Map will be enabled in the menu.