7-10
Operation Manual
Tips!
Biopsy Guide Edit Mode
NOTE:
X
Only available in 2D mode and Single mode.
X
Not available in Dual Live or Freeze state.
X
Not available in Write Zoom or Read Zoom state.
Starting and Ending Biopsy Guide Edit Mode
1. Tap Biopsy Guide on the touch screen.
2. Use the Angle dial-button on the control panel to enter Biopsy Guide Edit mode.
3. When you enter Biopsy Guide Edit mode, Angle Offset appears.
4. Rotate the Angle dial-button to select Angle Offset or Lateral Offset.
− Angle Offset: This is the angle for the currently selected biopsy guide; the maximum
adjustable offset is the guideline's outer line angle. Adjustments can be made in 0.1 degree
increments.
− Lateral Offset: This is the lateral direction for the currently selected biopsy guide. Adjustments
can be made in 0.1mm increments.
5. Press the Angle dial-button to exit Biopsy Guide Edit mode.
Spatial Compound (Optional)
NOTE: This item appears in the Flexible Soft menu only when a Linear Probe is used.
Tap Spatial Compound on the touch screen. The ‘SCI’ mark will be displayed in the image
information. Tap the button on the touch screen to select Spatial Compound Off, Low, Medium, or
High.
Trapezoidal
NOTE: The Trapezoidal item appears in the menu only when a Linear Probe is used.
Turn it on or off by tapping Trapezoidal on the touch screen.
In general, the rectangular frame provided by a Linear Probe is changed to a trapezoidal shape. This
allows a wider view of an image.