Chapter 8 Measurements and Calculations
Measurement Accuracy
Measurement values can vary depending on the nature of the ultrasound wave, the body’s response
to ultrasound waves, the measurement tools, algorithms, product settings, probe type and operations
performed by the user.
Before using this product, make sure to read and understand the following information regarding the
causes of measurement errors, and measurement optimization.
Causes of Measurement Errors
Image Resolution
The resolution of an ultrasound image may be limited by spatial causes.
X
Errors caused by signal range may be minimized by adjusting the focus settings. Optimizing
focus settings increases the resolution of the measurement area.
X
In general, lateral resolution is lower than axial resolution. Therefore, measurements should be
performed along the axis of the ultrasound beam to obtain accurate values.
X
Gain has a direct impact on resolution. Gain can be adjusted by using the Gain button for each
mode.
X
In general, increasing the frequency of ultrasound enhances resolution.
Pixel Size
X
This product’s ultrasound images consist of pixels.
X
Since a single pixel represents the basic unit of an image, a measurement error may result in the
displacement of approximately ±1 pixel when compared to the original image size.
X
However, this error is significant only when a narrow interval is being measured on the monitor.