9-12
Operation Manual
Tips!
e-Motion Marker (Optional)
If you are using e-Motion Marker, which is an optional feature for this product, you can run
e-Motion Marker by pressing the Body Marker button on the control panel. The e-Motion Marker
displays the direction of the probe’s beam plane on the screen.
WARNING:
X
The e-Motion Marker provides reference information to assist in diagnosis. It
does not provide any numerical diagrams or final diagnosis.
X
The e-Motion Marker displays the relative directions of a beam plane, so
the direction displayed may differ from the actual probe direction because
of accumulated errors. Therefore, press the Initialize button to initialize the
position before starting to use the e-Motion Marker.
NOTE:
X
e-Motion Marker can be used only if the following conditions are met:
− Probe: EVN4-9
− Application: Gynecology
− Preset: Uterus
X
Be sure to refer to the installation guide that comes with the e-Motion Marker
package before installing the e-Motion Marker.
X
How to use e-Motion Marker
Press the e-Motion Marker buttons to use their pre-assigned functions.
1: If pressed briefly, the Initialize button initializes the e-Motion Marker's sensor. Pressing and
holding the button for three seconds or longer will start Calibration mode. Pressing and holding
the button for three seconds or longer a second time will end Calibration mode. The user may
not assign another function to this button.
NOTE:
X
Since the strength of the magnetic field varies depending on the latitude of
your location, you need to perform a Sensor Calibration at least once when
you use the product for the first time.
X
For Sensor Calibration Settings, be sure to refer to the installation guide.
2~4: These buttons may be assigned functions as desired by the user. Go to Setup > Body
Marker > Body Marker > e-Motion Marker to assign a function to each button. Refer to ‘Body
Marker’ in ‘Chapter 3. Utilities’ for instructions on the setup.