8-80
Operation Manual
6. After one measurement is taken, the measurement item will return to the original undened
state. You may select the measurement item again before exiting the Auto IMT screen to continue
taking measurements.
Setting Up Measurement Items
In the left screen menu or on the touch screen, you may set up the measurement item, the location
of the blood vessel, and the direction of blood flow.
X
Measurement item: Select a measurement item from CCA, Bulb, ICA, and ECA. Use the left screen
menu, or use the touch screen button.
X
Direction: Select Left or Right as the direction of blood flow. Use the left screen menu, or the
dial-button 2.
X
Location: Select Prox (Proximal), Mid (Middle), or Dist (Distal) as the location of the blood vessel
to measure. Use the left screen menu, or the dial-button 3.
NOTE: Doing one of the following will exit Auto IMT:
X
Changing Image Mode, Zoom, or Depth
X
Annotation, BodyMarker, Patient, SonoView
Measurement Result
X
Max: The maximum thickness of the Intima/Adventitia pair.
X
Mean: The average thickness of the Intima/Adventitia pair
X
SD: Standard Deviation
X
QI: The distance ratio of the measured point in a distance for Quality Index measurement.
NOTE: You may click the measurement result window to reposition the window
Applying to the Report
The Auto IMT measurement result is automatically applied to the Report. Press the Vascular tab at
the left side of the Report screen to view Auto IMT measurement result.