1-23
Chapter 1 Safety
Additional Considerations
Ensure that scanning time is kept to a minimum, and ensure that only medically required scanning is
performed. Never compromise quality by rushing through an exam. A poor exam will require a follow-up,
which ultimately increases the time. Diagnostic ultrasound is an important tool in medicine, and, like any
tool, should be used eciently and eectively.
Output Display Features
The system output display comprises two basic indices: a mechanical index and a thermal index. The
thermal index consists of the following indices: soft tissue (TIs), bone (TIb) and cranial bone (TIc). One of
these three thermal indices will be displayed at all times. Which one depends upon the system preset or
user choice, depending upon the application at hand.
The mechanical index is continuously displayed over the range of 0.0 to 1.9, in increments of 0.1.
The thermal index consists of the three indices, and only one of these is displayed at any one time.
Each probe application has a default selection that is appropriate for that combination. The TIb or TIs is
continuously displayed over the range of 0.0 to maximum output, based on the probe and application, in
increments of 0.1.
The application-specific nature of the default setting is also an important factor of index behavior. A
default setting is a system control state which is preset by the manufacturer or the operator. The system
has default index settings for the probe application. The default settings are invoked automatically by the
ultrasound system when power is turned on, new patient data is entered into the system database, or a
change in application takes place.
The decision as to which of the three thermal indices to display should be based on the following criteria:
Appropriate index for the application: TIs is used for imaging soft tissue; and TIb for a focus at or near
bone. Some factors might create articially high or low thermal index readings e.g. presence of uid or
bone, or the ow of blood. A highly attenuating tissue path, for example, will cause the potential for local
zone heating to be less than the thermal index displays.
Scanned modes versus unscanned modes of operation affect the thermal index. For scanned modes,
heating tends to be near the surface; for unscanned modes, the potential for heating tends to be deeper
in the focal zone.
Always limit ultrasound exposure time. Do not rush the exam. Ensure that the indices are kept to a
minimum and that exposure time is limited without compromising diagnostic sensitivity.
Vista de página 38
1 2 ... 34 35 36 37 38 39 40 41 42 43 44 ... 267 268

Comentários a estes Manuais

Prof Akhigbe 24 Apr 2024 | 14:29:06

I want to learn how to do angle correction on PW Doppler on my machine.

Prof Akhigbe 24 Apr 2024 | 14:30:30

How do I assess PW angle cursor?

Prof Akhigbe 24 Apr 2024 | 14:40:21

I don't want to join the group.