5-7
Chapter 5 Measurements and Calculations
Measurement Accuracy Table
The following tables show the accuracy of the measurements available using the product. Ensure that the
results of measurement accuracy checks are kept within the ranges specied in the table. Except for certain
applications or probes, the following accuracy ranges should be maintained for measurement of a straight
distance.
NOTE:
To ensure accurate measurements, an accuracy check should be performed at least once per year.
If the measurement accuracy falls outside the ranges specified in the following table, contact Samsung
Medison Customer Service.
2D Mode
Measurements
System Tolerance
(Whichever is greater)
Test Methodology
Accuracy
(1)
Based on
Range
(2)
Axial Distance < +/- 4% or 1mm Phantom Acquisition Full Screen
Lateral Distance < +/- 4% or 2mm Phantom Acquisition Full Screen
Axial Resolution < = 2mm Phantom Acquisition Full Screen
Lateral Resolution < = 3mm Phantom Acquisition Full Screen
(1) The accuracy of the measurements diers from the table above, depending on the user’s skills.
(2) The distance is indicated as cm’ up to two decimal points.
M Mode
Measurements
System Tolerance
(whichever is greater)
Test
Methodology
Accuracy
(1)
Based on
Range
(2)
Depth <+/- 5% or 3 mm Phantom Phantom 1 - 25 cm
Time < +/- 5%. Signal generator Phantom .01 - 11.3 sec
(1) The accuracy of the measurements diers from the table above, depending on the user’s skills.
(2) The distance is indicated as cm’ up to two decimal points. The speed is indicated as cm/s’ up to two
decimal points.
Vista de página 168
1 2 ... 164 165 166 167 168 169 170 171 172 173 174 ... 267 268

Comentários a estes Manuais

Prof Akhigbe 24 Apr 2024 | 14:29:06

I want to learn how to do angle correction on PW Doppler on my machine.

Prof Akhigbe 24 Apr 2024 | 14:30:30

How do I assess PW angle cursor?

Prof Akhigbe 24 Apr 2024 | 14:40:21

I don't want to join the group.