4-7
Chapter 4 Diagnosis Modes
2D Mode Soft Menu
The items that are commonly used in 2D Mode during scanning are provided as Soft Menu items.
You can use the Soft Menu dial-button corresponding to a specific menu item on the control panel.
If you rotate the dial-button, the setting for the selected item changes.
Dynamic Range / Harmonic
Use the Soft Menu dial-button [1].
Dynamic Range
The contrast of an image is adjusted by adjusting the ratio of the minimum and maximum values
of the input signals. Select a value between 50 and 200 by rotating the Soft Menu dial-button [1].
The larger the ratio, the smoother the image gets.
Harmonic
When pressing the Soft Menu dial-button [1], the HAR indication appears on the image
information. The OHI (Optimal Harmonic Imaging) function, which allows you to optimize images
by using high frequencies, is provided.
NOTE:
The Harmonic function can be used with specific probes.
Reject Level / 2D/C Live
Use Soft Menu dial-button [2].
Reject Level
Noise or echo is removed to make the image clearer. Select a value between 1 and 32 by rotating
the Soft Menu dial-button [2].
2D/C Live
Press the Soft Menu dial-button [2]. You can observe a scanned area in 2D and Color Doppler
images simultaneously in real time.
FSI / M Line
Use the Soft Menu dial-button [3].
Vista de página 104
1 2 ... 100 101 102 103 104 105 106 107 108 109 110 ... 267 268

Comentários a estes Manuais

Prof Akhigbe 24 Apr 2024 | 14:29:06

I want to learn how to do angle correction on PW Doppler on my machine.

Prof Akhigbe 24 Apr 2024 | 14:30:30

How do I assess PW angle cursor?

Prof Akhigbe 24 Apr 2024 | 14:40:21

I don't want to join the group.