4-10
Operation Manual
■
Gray Map / L/R Flip
▶
Gray Map
Changes a 2D Post Curve. Select from Type 1 to Type 9 or User 1 to User 2 by rotating the Soft
Menu dial-button [2].
▶
L/R Flip
When the Soft Menu dial-button [2] is pressed, the left and right sides of an image are ipped.
The M indication at the top of the image indicates the direction of the current image.
■
Focus / U/D Flip
▶
Focus
Use the Soft Menu dial-button [3] to set the number of focusing points to between 1 and 4.
▶
U/D Flip
Press the Soft Menu dial-button [3]. Whenever the dial-button is pressed, an image is flipped
upside down.
■
Sector Width
Change the sector width. Select a value of between 40 to 100 % by rotating the Soft Menu dial-button
[4]. When the width is increased, the frame rate is reduced.
NOTE:
If the application is set as Fetal Heart, you can select up to 20%.
■
Frame Avg
When images are updated, the previous and current images are averaged. Select a value between 0
and 15 by rotating the Soft Menu dial-button [5]. If you scan the same area repeatedly, speckles may
appear on the updated image. This item is used to minimize this phenomenon.
■
Power
Adjusts the intensity of the ultrasound output. Select a value between 10 and 100 by rotating the Soft
Menu dial-button [6].