4-40
Operation Manual
Load Preset
Select a preset for the 3D image. Select Default or User1 through User5 by rotating the soft menu’s
dial-button [3]. For more information, refer to the preset information contained in the ‘3D View-MPR
Mode’ section.
Scan Quality
Set the 3D image quality. Select from Low, Medium, High, and Extreme by rotating the soft menu’s
dial-button [5].
X
Extreme: Provides superior image quality. Use for studying a highly detailed image.
X
High: Provides higher 3D image capturing (or rendering) speed than Extreme setting, at the
expense of image quality.
X
Middle: Provides better image capturing speed and lower image quality than High setting.
X
Low: Provides the highest 3D image capturing speed and the lowest image quality.
Auto ROI
NOTE: Auto ROI only appears in the menu when the 3D probe’s application is OB.
Press the soft menu’s dial-button [5] to turn Auto ROI on or o. Once turned on, the volume box is
automatically positioned over the area that will be converted into 3D.
Tips!
Auto ROI Considerations
When Auto ROI is used, the following should be taken into account:
X
Only images of the fetal body can be acquired.
X
Auto ROI is affected by the brightness or contrast of 2D images.
X
Once Auto ROI is turned on, the position or size of the volume box cannot be changed.
Scan Angle
Used to adjust the scan angle. Rotate the soft menu’s dial-button [6] to adjust the scan angle. The
available scan angle range varies depending on the probe being used.