Samsung MYSONO U6 Manual do Utilizador Página 132

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 282
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 131
4-24
Operation Manual
Starting & Ending PW Spectral Doppler Mode
Press the PW button on the control panel to enter PW Spectral Doppler mode. Press it again to return
to 2D Mode.
Press the Set button on the control panel to obtain a spectral doppler image.
PW Spectral Doppler Mode Screen
Sample Volume
The Doppler spectrum is displayed when the sample volume is located above the blood ow on the
2D image. The size and depth of Sample volume is displayed in mm units. Its position is moved with
the trackball and displayed in the xx.xx@yy.yy mm format. The information is formatted to indicate
a sample volume that is xx.xx mm in size at a depth of yy.yy mm.
For example, [email protected] means a 2.00mm sample volume at 16.70mm in depth.
X
Moving Sample Volume: Use the trackball on the control panel.
X
Resizing Sample Volume After pressing the Change button on the control panel, adjust the size
of the sample volume by using the trackball. Press the Change button again to return to the
Sample Volume Position Control screen. Here, an icon is displayed in the upper left corner of the
screen to indicate whether the trackball is set to SV Position or SV Size.
X
Adjusting Sample Volume Angle: Use the control panel’s Angle dial-button to adjust the sample
volume angle. Rotating the Angle dial-button clockwise increases the angle within a range of
–70 to +70.
HPRF
NOTE:
X
Enable/disable HPRF from Utility > Setup > General > Scan Mode > Option.
X
HPRF is not activated in Simultaneous Mode. In addition, HPRF cannot be enabled if 200% of PRF
is higher than 23KHz.
Adjust the blood ow above the speed limit at a specic depth. Using this functions doubles the
current scale, and is only available in PW Spectral Doppler mode. HPRF stands for High PRF.
X
Activating HPRF
Increasing the scale at a specic depth to a certain point automatically activates HPRF. A The
Phantom Gate will appear on the D Line at a position higher than the sample volume. Once HPRF
starts, PRF does not increase even if you increase the scale value.
Vista de página 131
1 2 ... 127 128 129 130 131 132 133 134 135 136 137 ... 281 282

Comentários a estes Manuais

Sem comentários