4-36
Operation Manual
2D/C/M Mode
Color Doppler Mode and M Mode are displayed simultaneously.
In Color Doppler Mode, press the M button on the control panel. Or, in M mode, press the C button on
the control panel. (This mode is enabled for specic diagnostic applications with specic probes only.)
2D/C Live Mode
2D Mode and Color Doppler Mode are displayed simultaneously. In 2D mode, select the soft menu’s dial-
button [5] 2D/C Live.
Changing Combined Mode Format
Changing the Active Image Mode
Press the Set button on the control panel. The current active image mode – ‘D Only’ or ‘2D Only’ - is
displayed above the menu on the screen.
In Combined Mode, more than two image modes are used at the same time. The image mode
currently in use is called the “Active Image Mode”. For example, if Sample Volume is moved with the
trackball in 2D/C/PW Mode, PW Spectral Doppler Mode becomes the current active image mode.
Because the menu and button options vary depending on the active image mode, you can use the
Set button to change the active image mode.
Note that the active image mode cannot be changed with the Set button when the Freeze function
is in eect.
Changing the menu
Press the Active Mode button on the control panel.
This changes the menu and the soft menu without aecting the active image mode. The functions
of the buttons on the control panel vary depending on the active image mode.
For example, pressing the Active Mode button while in 2D/C/PW mode with the 2D mode soft
menu loaded on the screen loads up Color Doppler mode’s soft menu.
NOTE: For information on optimizing an image in Combined Mode, please refer to ‘Basic Mode’.