4-10
Operation Manual
U/D Flip
Press the soft menu’s dial-button [2]. Each press of the dial-button ips the image vertically.
Line Density
Set the scan line density. Select from Medium, High, and Low by rotating the [3] dial button beneath
the soft menu.
Selecting High increases the number of scan lines and improves the image resolution. However, the
frame rate decreases in this case.
Trapezoidal
NOTE: Trapezoidal appears in the menu only when a linear probe is used.
Press the soft menu’s dial-button [3] to turn this option on or o. In general, the rectangular frame
provided by a linear probe is changed to a trapezoidal frame when this option is turned on. This
allows a wider view of an image.
The Trapezoidal function may not be available for certain depths. In addition, the Zoom button
cannot be used while the Trapezoidal function is being used.
Pulse Inversion
Press the soft menu’s dial-button [4] to turn Pulse Inversion on or o. If it is turned on, pulses are
inverted to sharpen the displayed image.
NOTE: Pulse Inversion is only available with specific probes.
Power
Used to adjust the ultrasound output. Rotate the soft menu’s dial-button [5] and set to a value
between 10 and 100.
Tissue
Optimizes the image by selecting the ideal frequency and speed settings for the type of tissue
being scanned. Rotate the soft menus dial-button [5] and choose from Solid, Normal, Adipose,
and Cystic.