Color Invert
Pressing the soft menu’s dial-button [1] inverts the color bar. Inverting the color bar also inverts
the color displayed on the image.
Color Map
Used to congure the post curve in a color image. Choose from Type 1 through 14 by rotating the
[3] dial-button beneath the soft menu.
Sensitivity
Used to adjust the color image sensitivity. Rotate the soft menu’s dial-button [4] and set to a value
between 8 and 31. When this value increases, the sensitivity of a color image improves, but the
frame rate becomes lower.
Balance
The range of a color image is adjusted by comparing the gray levels of 2D images with the
doppler signal values of color images. Rotate the soft menu’s dial-button [5] and set to a value
between 1 and 16.
When the Balance value increases, the color image also appears in the part where the gray level of
a 2D image is high (the bright part), increasing the range of the color image.
Scale
Use the soft menu’s dial-button [6] to adjust scale. Rotating the dial-button in the clockwise
direction increases the PRF (Pulse Repetition Frequency), displaying a wider range of blood ow
speed. Rotating the dial-button in the counterclockwise direction decreases the PRF, displaying a
narrower range of blood ow speed.
Color Format
Select C mode to use. Rotate the soft menu’s dial-button [1] and select between Color + B/W and
B/W Only.
Color Mode
Used to specify the display method in Color Doppler mode. Rotate the soft menu’s dial-button
[2] and select between Velocity and Vel + Var. Vel is an abbreviated form of velocity, and Var is an
abbreviated form of variance.