Samsung MYSONO U6 Manual do Utilizador Página 115

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 282
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 114
4-7
Chapter 4 Diagnosis Modes
2D Mode Soft Menu
The soft menu in 2D mode consists of the following:
NOTE: The soft menu referred in this manual can be accessed from default settings at Utility >
Menu Edit.
1 2 3 4 5 6
Frame Avg Dynamic Range Reject Level Edge Enhance Focus DMR+ Index
Frequency Harmonic M Line Write Zoom 2D/C Live DMR+
Rotation Frame Rate Line Density Power SRF Index
L/R Flip U/D Flip Trapezoidal Pulse Inversion Tissue SRF
Frame Avg
When an image is updated, the present image and the previous image are averaged. Set Frame
Avg to between 0 and 15 by rotating the [1] dial button beneath the soft menu. When you scan the
same diagnosis site repeatedly, speckles may appear on the updated image. Frame Avg is used to
minimize these speckles.
Frequency
Used to change the probes frequency. Select from Res, Pen, and Gen by rotating the [1] dial button
beneath the soft menu.
X
Res (Resolution): High frequency
X
Gen (General): General frequency
X
Pen (Penetration): Low frequency
The selected frequency is displayed in the title area, allowing you to determine the state of the
current frequency easily.
Dynamic Range
Adjust the image contrast by changing the minimum and maximum values of the input signal. Turn
the soft menu’s dial-button [2] and select a value between 50 and 200. If you increase the range, the
image will become smoother.