GE HEALTHCARE
DIRECTION 5194296-100, REVISION A LOGIQ BOOK XP SERIES BASIC SERVICE MANUAL
Chapter 10 Care & Maintenance 10-15
10-6-4-3 Data Sheet for enclosure Source Leakage Current
The test passes when all readings measure less than the value shown in Table 10-11. Record all data
on the PM Inspection Certificate.
Table 10-14 Typical Data Sheet for enclosure Source Leakage Current
Unit Power
Tester Polarity
Switch
Tester Neutral
or Ground
Switch
Test 1
Speaker
Cover
Test 2
Real Panel
Metal
Parts
Optional
Test 3
Optional
Test 4
Enter Name of tested peripheral here:
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
OFF NORM OPEN
OFF NORM CLOSED
OFF REV OPEN
OFF REV CLOSED
Comentários a estes Manuais