
GE HEALTHCARE
D
IRECTION 5194296-100, REVISION 10 LOGIQ BOOK XP SERIES BASIC SERVICE MANUAL
Chapter 4 Functional Checks 4-13
4-3-8 M Mode Controls
4-3-8-1 Preparations
1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-24, in Chapter 3 Installation
to the System probe connector.
2.) Turn ON the system (if it isn’t turned on already).
Figure 4-11 Controls available in M Mode
Figure 4-12 M Mode Screen Picture Example
Primary Menu
secondary Menu
M Mode
Gray Map
Dynamic Range
Sweep Speed
Display Format
Full Timeline
Colorize
Edge Enhance
Rejection
Power Output
Gain
M/D Cursor
Comentários a estes Manuais