GE HEALTHCARE
D
IRECTION 5194296-100, REVISION 10 LOGIQ BOOK XP SERIES BASIC SERVICE MANUAL
Chapter 4 Functional Checks 4-5
4-3-3 Archiving and Loading Presets
NOTE: Always save presets before any software reload. This ensures the presets loaded after the software
reload are as up–to–date as possible.
All user presets except changes to Summary, Anatomy, and Biometry pages, can be saved on an CD-
R disk for reloading on the system.
4-3-3-1 Archiving Presets to an CD-R disk
1.) Insert an empty (blank) CD-R disk into the CD-RW.
2.) Access to the Utility Menu, and select Backup/Restore. The Backup sheet will be shown on the
monitor.
3.) Select the item to back up either from Resource Files.
4.) Enter backup destination or browse through the disk to locate the destination.
5.) Select Backup now. The backup status for each item is displayed on the Result column.
4-3-3-2 Loading Presets from an CD-R disk
1.) Insert the CD-R disk with the archived Presets into the CD-RW.
2.) Access to the Utility Menu, and select Backup/Restore. The Restore sheet will be shown on the
LCD display. (see Figure 4-4 on page 4-5 )
3.) Select the item to restore either from Resource Files.
4.) Enter restore destination or browse through the disk to locate the destination.
5.) Select Restore. The restore status for each item is displayed on the Result column.
NOTICE
Presets should NOT be saved on the same CD-R disk as images. The Archive Menu lists the images
but does NOT list the presets stored on a CD-R disk.
Figure 4-4 Backup Sheet