Chapter 5.Measurements and Calculations 5 - 5
Optimization of Measurement Accuracy
2D Mode
- Resolution is in proportion to the frequency of the probe.
- Penetration is in inverse proportion to the frequency of the probe.
- The highest resolution can be obtained at the focus of the probe where the ultrasound
beam is narrowest.
- The most accurate measurements can be obtained at the focus depth. The accuracy
decreases as the distance from the focus increases, widening the beam width.
- Using the zoom function or minimizing the depth display makes distance or area
measurements more accurate.
M Mode
- The accuracy of time measurements can be increased when the sweep velocity and the
display format are set to high values.
- The accuracy of distance measurements can be increased when the display format is
set to higher values.
Doppler Mode
- It is recommended to use lower frequency ultrasound for measurement of faster blood
flows.
- The size of the sample volume is limited by the axial direction of the ultrasound.
- Using lower frequency ultrasound increases penetration.
- The accuracy of time measurements can be increased when the sweep velocity is
increased.
- The accuracy of velocity measurements can be increased when the vertical scale is set
to smaller values.
- It is most important to use an optimal Doppler angle to enhance the accuracy of velocity
measurements.
Color/Power Doppler Mode
- A protocol is not specified for images in Color Doppler Mode or Power Doppler Mode.
Therefore, the same limitations imposed when measurements are taken in B/W images
apply to the accuracy of the measurements taken in these modes.