Chapter 5.Measurements and Calculations 5 - 15
specify a point in a Spectral Doppler image and trace a curve from that point to calculate the
velocity, integral value and average velocity of blood flow.
1. Select the Flexible Soft button [1] D Trace. “D Trace” is displayed in the user
information area.
2. Trace a curve. The method for measuring a curve is the same as in “2D Line Trace.”
3. When the measurement is finished, its result is shown on the screen.
- Vmax: Max. Velocity
- Vmean: Average Velocity
- PGmax
- PGmean
- VTI: VTI (Velocity Time Integral)
- Time: Time Change
4. A new measurement cursor appears.
The equations used for D Trace measurement are as follows:
-
12
TTdT =
-
):(, IntegralTimeVelocityVTIdtvVTI
×=
-
flowofDuration
VTI
meanV =