Chapter 5.Measurements and Calculations 5 - 45
Vascular Measurements
Menu Sub menu Input Meas.
Rt./ Lt. Subclavian
Trace Auto, Limited, Manual Doppler Waveform Trace
PSV Velocity
EDV Velocity
%StA Calculation
%StD Calculation
Vessel Area Area
Vessel Distance Distance
Rt./Lt. Prox CCA
(Proximal Common Carotid Artery)
Same as above
Same as above
Rt./Lt. Mid CCA Same as above Same as above
Rt./Lt. Distal CCA Same as above Same as above
Rt./Lt. Bulb Same as above Same as above
Rt./Lt. Prox ICA
(Proximal Internal Carotid Artery)
Same as above
Same as above
Rt./Lt. Mid ICA Same as above Same as above
Rt./Lt. Distal ICA Same as above Same as above
Rt./Lt. ECA
(External Carotid Artery)
Same as above
Same as above
Rt./Lt. Vertebral Same as above Same as above
General Same as above Same as above
The values such as PSV, EDV, Vmean, Gmean, Gpeak, S/D, PI, RI, etc. can be
obtained from the spectral waveform in Spectral Doppler Mode. There are four
measurement methods: Auto Trace, limited Trace, Manual Trace, and Measurement
Each Item One by One. For more information, refer to Doppler Spectrum
Measurement Method.
Stenosis % Calculation : There are two ways to obtain the Stenosis % of each
vascular system: %StA using the area and %StD using the distance. Refer to Volume
Flow Measurementfor more information.
Vessel Area Measurement: There are two ways to obtain Vessel Area: Vesl. Area
using the circumference and Vesl. Dist using the distance.
The user can obtain the Volume Flow values using the TAM (Time Average Mean
Velocity) value obtained from the spectral waveform and the Vessel Area value
obtained using Vesl. Area or Vesl. Dist.