Chapter 5.Measurements and Calculations 5 - 13
Please refer to the table below for Hip Joint Type information:
Type α β
1a
60 ≤ α< 90 0 < β< 55
1b
60 ≤ α< 90 55 ≤ β< 90
2a/b
50 ≤ α< 60 0 < β< 90
2c
43 ≤ α< 50 77 ≤ β< 90
d
43 ≤ α< 50 0 < β< 77
3/4
0 < α< 43
[Table 5.2 Hip Joint Type Table]
▐ M Distance
This is a basic measurement that is available in M Mode only. You can specify two points in
an M image and measure the distance, elapsed time and velocity between them.
1. Select the Flexible Soft button [1] M Distance. “M Distance” is displayed in the user
information area.
2. Specify two points and measure the straight distance between them. The method for
measuring a line is the same as in “2D Distance.”
3. When the measurement is finished, its result is shown on the screen.
- The measured distances are shown as “Dist.” and can be differentiated with their
measurement cursor.
- The elapsed time (Time) and velocity (Slope) are also displayed.
4. A new measurement cursor appears. When more than 4 measurements are taken,
measurements are deleted in the order in which they have been taken.
▐ Doppler Velocity
This is a basic measurement that is available in Spectral Doppler Mode only. You can
specify two points in a Spectral Doppler image and measure the distance between them,
and the velocity at each point to calculate the velocity change, time change and acceleration.
NOTE
In a Spectral Doppler image, the X- and Y-axes represent time and
velocity, respectively.
You can measure acceleration for up to two segments and differentiate the points of
segments into Point 1 and 2.
1. Select the Flexible Soft menu [1] D Velocity. “D Velocity” is displayed in the user
information area.
Comentários a estes Manuais