Samsung SONOACE R7 Manual do Utilizador Página 27

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 320
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 26
1-11
Chapter 1 Safety
EMI
Although this system has been manufactured in compliance with existing EMI (Electromagnetic Interference)
requirements, use of this system in the presence of an electromagnetic eld can cause momentary degradation
of the ultrasound image.
If this occurs often, Samsung Medison suggests a review of the environment in which the system is being used,
to identify possible sources of radiated emissions. These emissions could be from other electrical devices used
within the same room or an adjacent room. Communication devices such as cellular phones and pagers can
cause these emissions. The existence of radios, TVs, or microwave transmission equipment nearby can also
cause interference.
CAUTION: In cases where EMI is causing disturbances, it may be necessary to relocate this system.
EMC
The testing for EMC (Electromagnetic Compatibility) of this system has been performed according to the
international standard for EMC with medical devices (IEC 60601-1-2). This IEC standard was adopted in Europe
as the European norm (EN 60601-1-2).
Guidance and manufacturer’s declaration - electromagnetic emission
This product is intended for use in the electromagnetic environment specied below. The customer or the
user of this product should assure that it is used in such an environment.
Emission test Compliance ElectroMagnetic environment -guidance
RF Emission
CISPR 11
Group 1
The Ultrasound System uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
RF Emission
CISPR 11
Class B
The Ultrasound System is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
building used for domestic purpose.
Harmonic Emission
IEC 61000-3-2
Class A
Flicker Emission
IEC 61000-3-3
Complies