5-12
Operation Manual
Angle
A basic measurement that is available in all diagnosis modes. Specify two straight lines in a 2D image and
measure the angle between them.
1.
Select
Angle
in the
Soft Menu
[1]
. Angle will appear in the user information area.
2.
Draw two straight lines. For information on drawing a straight line, please refer to ‘Distance.
3.
The angle between two lines will be calculated and displayed on the screen.
X When two angles are calculated, the smaller angle is displayed.
%StD
StD stands for Stenosis Distance, which is a basic measurement available in all diagnosis modes. In a 2D
image, the diameter of a vessel is measured and the stenosis ratio calculated.
1.
Sele
ct %StD in
the
Soft Menu
[1].
%StD will a
ppear in the user information area.
2.
Measure the total diameter of a vessel using the Distance measurement method.
3.
When a new cursor appears, measure the inner wall diameter of the vessel under stenosis.
4.
Calculate %StD with the following equation:
X
%StD = (Outer Distance – Inner Distance) / Outer Distance × 100
M Distance
This is a basic measurement that is available in M Mode only. You can specify two points in an M image and
measure the distance, elapsed time and velocity between them.
1.
Sele
ct M Distance in the
Soft Menu
[1].
M Distance will appear in the user information area.
2.
Specify two points and then measure the shortest distance between them. The way that it is measured
is the same as in ‘Distance.
3.
Once measurement is done, the result is displayed on the screen.