Samsung SONOACE R7 Manual do Utilizador Página 178

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 320
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 177
5-4
Operation Manual
Doppler Signal Adjustment
X During velocity measurement, an error may occur depending on the cosine angle between the blood
ow and the ultrasound beam.
X For Doppler velocity measurements, the most accurate results can be ensured when the ultrasound
beam is aligned in parallel with the blood ow.
X If that is not possible, the angle between them should be adjusted by using the
Angle
option.
Aliasing
X PW Spectral Doppler Mode uses a signal sampling technique to calculate the frequency (or velocity)
spectrum.
X Adjust the baseline or the velocity scale to minimize aliasing. A lower frequency probe can also be
used to reduce aliasing.
X Aliasing is dramatically reduced in CW Spectral Doppler Mode.
Calculation Equation
X Some of the calculation equations used for clinical purposes originate from hypotheses and
approximation.
X All calculation equations are based on medical reports and articles.
Human Error
X Human error may occur due to inappropriate use or lack of experience.
X This can be minimized through compliance with and thorough understanding of the manuals.
Vista de página 177
1 2 ... 173 174 175 176 177 178 179 180 181 182 183 ... 319 320

Comentários a estes Manuais

Sem comentários