Operation Manual
Electrical Safety
This product is categorized as a Class I device.
CAUTION:
X
As for US requirement, the LEAKAGE CURRENT might be measured from a center-tapped circuit
when the equipment connects in the United States to 240V supply system.
X
To help assure grounding reliability, connect to a “hospital grade” or “hospital only” grounded
power outlet.
Prevention of Electric Shock
In a hospital, dangerous currents are due to the potential dierences between connected equipment and
touchable conducting parts found in medical rooms. The solution to the problem is consistent equipotential
bonding. Medical equipment is connected with connecting leads made up of angled sockets to the
equipotential bonding network in medical rooms.
[Figure 1.1 Equipotential bonding]